Application of Bioabsorbable Plates in Orthognathic Surgery

  • Kim, Young-Kyun (Department of Oral and Maxillofacial Surgery, Section of Dentistry, Seoul National University Bundang Hospital) ;
  • Shim, Cheong-Hwan (Department of Oral and Maxillofacial Surgery, Section of Dentistry, Seoul National University Bundang Hospital) ;
  • Bae, Ji-Hyun (Department of Conservative Dentistry, Section of Dentistry, Seoul National University Bundang Hospital) ;
  • Yun, Pil-Young (Department of Oral and Maxillofacial Surgery, Section of Dentistry, Seoul National University Bundang Hospital)
  • 발행 : 2006.02.28

초록

Backgrounds: Though they are considered as reliable devices, titanium plates and screws have limitations due to some potential problems. To overcome these problems, researches on bioabsorbable materials for internal fixation have been continuing. Recently, there are many clinical trials to apply biodegradable internal fixation devices in oral and maxillofacial surgery. The purpose of this study was to evaluate the clinical results of application of bioabsorbable plates and screws in orthognathic surgery. Methods: Fifty-four patients with dentofacial deformity (17 male patients and 37 female patients) were selected in this study. The patients were treated by internal fixation using bioabsorbable plates and 2.4-mm, 2.0-mm pretapped screws (Biosorb FX, Bionix Implants, Inc., Finland). The patients were evaluated for complications during the follow-up period. Results: Five patients (9.3%) experienced complications. All complications in these cases were infection. No other complications related with physical or mechanical properties of bioabsorbable plates were found such as malunion or nonunion, fractures of plates and loosening of screws. All complications were minor and adequately managed with drainage and supportive care with antibiotics coverage. Conclusions: From the results, the use of these fixation systems in orthognathic surgery will provide a promising alternative titanium fixation in appropriate cases.

키워드

참고문헌

  1. Tuovinen V, Norholt SE, Sindet-Pedersen S, Jensen J: A retrospective analysis of 279 patients with isolated mandibular fractures treated with titanium miniplates. J Oral Maxillofac Surg 1994;52:931-93587 https://doi.org/10.1016/S0278-2391(10)80072-2
  2. Matthew IR, Frame JW: Policy of consultant oral and maxillofacial surgeons towards removal of miniplate components after jaw fracture fixation: pilot study. Br J Oral Maxillofac Surg 1999;37:110-112 https://doi.org/10.1054/bjom.1997.0084
  3. Iizuka T, Lindqvist C: Rigid internal fixation of mandibular fractures. An analysis of 270 fractures treated using the AO/ASIF method. Int J Oral Maxillofac Surg 1992;21:65-69 https://doi.org/10.1016/S0901-5027(05)80533-8
  4. Fiala TG, Novelline RA, Yaremchuk MJ: Comparison of CT imaging artifacts from craniomaxillofacial internal fixation devices. Plast Reconstr Surg 1993;92:1227-1232
  5. Fiala TG, Paige KT, Davis TL, Campbell TA, Rosen BR, Yaremchuk MJ: Comparison of artifact from craniomaxillofacial internal fixation devices: magnetic resonance imaging. Plast Reconstr Surg 1994;93:725-731 https://doi.org/10.1097/00006534-199404000-00011
  6. Schliephake H, Reiss G, Urban R, Neukam FW, Guckel S: Metal release from titanium fixtures during placement in the mandible: an experimental study. Int J Oral Maxillofac Implants 1993;8:502-511
  7. Jorgenson DS, Mayer MH, Ellenbogen RG, Centeno JA, Johnson FB, Mullick FG, Manson PN: Detection of titanium in human tissues after craniofacial surgery. Plast Reconstr Surg 1997;99:976-979 https://doi.org/10.1097/00006534-199704000-00006
  8. Eppley BL, Sadove AM, Havlik RJ: Resorbable plate fixation in pediatric craniofacial surgery. Plast Reconstr Surg 1997;100:1-7 https://doi.org/10.1097/00006534-199707000-00001
  9. Tharanon W, Sinn DP, Hobar PC, Sklar FH, Salomon J: Surgical outcomes using bioabsorbable plating systems in pediatric craniofacial surgery. J Craniofac Surg 1998;9:441-444 https://doi.org/10.1097/00001665-199809000-00008
  10. Ferretti C, Reyneke JP: Mandibular, sagittal split osteotomies fixed with biodegradable or titanium screws: a prospective, comparative study of postoperative stability. Oral Surg Oral Med Oral Pathol Oral Radiol Endod 2002; 93:534-537 https://doi.org/10.1067/moe.2002.124091
  11. Kim YK, Yeo HH, Lim SC: Tissue response to titanium plates: a transmitted electron microscopic study. J Oral Maxillofac Surg 1997;55:322-326 https://doi.org/10.1016/S0278-2391(97)90115-4
  12. Cutright DE, Hunsuck EE, Beasley JD: Fracture reduction using a biodegradable material, polylactic acid. J Oral Surg 1971;29:393-397
  13. Suuronen R: Biodegradable fracture-fixation devices in maxillofacial surgery. Int J Oral Maxillofac Surg 1993;22:50-57 https://doi.org/10.1016/S0901-5027(05)80358-3
  14. Suuronen R, Pohjonen T, Vasenius J, Vainionpaa S: Comparison of absorbable self-reinforced multilayer polyl- lactide and metallic plates for the fixation of mandibular body osteotomies: an experimental study in sheep. J Oral Maxillofac Surg 1992;50:255-262 https://doi.org/10.1016/0278-2391(92)90322-Q
  15. Kallela I, Laine P, Suuronen R, Iizuka T, Pirinen S, Lindqvist C: Skeletal stability following mandibular advancement and rigid fixation with polylactide biodegradable screws. Int J Oral Maxillofac Surg 1998;27:3-8 https://doi.org/10.1016/S0901-5027(98)80086-6
  16. Shand JM, Heggie AA: Use of a resorbable fixation system in orthognathic surgery. Br J Oral Maxillofac Surg 2000; 38:335-337 https://doi.org/10.1054/bjom.1999.0287
  17. Turvey TA, Bell RB, Tejera TJ, Proffit WR: The use of selfreinforced biodegradable bone plates and screws in orthognathic surgery. J Oral Maxillofac Surg 2002;60:59-65 https://doi.org/10.1053/joms.2002.28274
  18. Haers PE, Sailer HF: Biodegradable self-reinforced poly- L/DL-lactide plates and screws in bimaxillary orthognathic surgery: short term skeletal stability and material related failures. J Craniomaxillofac Surg 1998;26:363-372 https://doi.org/10.1016/S1010-5182(98)80069-3
  19. Schmidt BL, Perrott DH, Mahan D, Kearns G: The removal of plates and screws after Le Fort I osteotomy. J Oral Maxillofac Surg 1998;56:184-188 https://doi.org/10.1016/S0278-2391(98)90865-5