A Study on the Effectiveness of the Manufacture of Compensator and Setup Position for Total Body Irradiation Using Computed Tomography-simulator's Images

전산화 단층 모의치료기(Computed Tomography Simulator)의 영상을 이용한 TBI(Total Body Irradiation) 자세 잡이 및 보상체 제작의 유용성에 관한 고찰

  • Lee Woo-Suk (Department of Radiation Oncology, Asan Medical Center, University of Ulsan) ;
  • Park Seong-Ho (Department of Radiation Oncology, Asan Medical Center, University of Ulsan) ;
  • Yun In-Ha (Department of Radiation Oncology, Asan Medical Center, University of Ulsan) ;
  • Back Geum-Mun (Department of Radiation Oncology, Asan Medical Center, University of Ulsan) ;
  • Kim Jeong-Man (Department of Radiation Oncology, Asan Medical Center, University of Ulsan) ;
  • Kim Dae-Sup (Department of Radiation Oncology, Asan Medical Center, University of Ulsan)
  • 이우석 (울산대학교의과대학 서울아산병원 방사선종양학과) ;
  • 박성호 (울산대학교의과대학 서울아산병원 방사선종양학과) ;
  • 윤인하 (울산대학교의과대학 서울아산병원 방사선종양학과) ;
  • 백금문 (울산대학교의과대학 서울아산병원 방사선종양학과) ;
  • 김정만 (울산대학교의과대학 서울아산병원 방사선종양학과) ;
  • 김대섭 (울산대학교의과대학 서울아산병원 방사선종양학과)
  • Published : 2005.09.01

Abstract

Purpose : We should use a computed tomography-simulator for the body measure and compensator manufacture process was practiced with TBI's positioning in process and to estimate the availability.,Materials and Methods : Patient took position that lied down. and got picture through computed tomography-simulator. This picture transmitted to Somavision and measured about body measure point on the picture. Measurement was done with skin, and used the image to use measure the image about lungs. We decided thickness of compensator through value that was measured by the image. Also, We decided and confirmed position of compensator through image. Finally, We measured dosage with TLD in the treatment department.,Results : About thickness at body measure point. we could find difference of $1{\sim}2$ cm relationship general measure and image measure. General measure and image measure of body length was seen difference of $3{\sim}4$ cm. Also, we could paint first drawing of compensator through the image. The value of dose measurement used TLD on head, neck, axilla, chest(lungs inclusion), knee region were measured by $92{\sim}98%$ and abdomen, pelvis, inquinal region, feet region were measured by $102{\sim}109%$.,Conclusion : It was useful for TBI's positioning to use an image of computed tomography-simulator in the process. There was not that is difference of body thickness measure point, but measure about length was achieved definitely. Like this, manufacture of various compensator that consider body density if use image is available. Positioning of compensator could be done exactly. and produce easily without shape of compensator is courted Positioning in the treatment department could shortened overall $15\{sim}20$ minute time. and reduce compensator manufacture time about 15 minutes.

목 적 : TBI의 자세 잡이 과정에서 실시하는 신체 계측과 보상체 제작 과정을 전산화 단층 모의치료기를 이용하고 그 유용성을 평가하고자 한다. 대상 및 방법 : 환자는 누운 자세를 취하고, 전산화 단층 모의치료기를 통해 영상을 얻었다. 이 영상을 Somavision으로 전송하고 영상을 통하여 신체 계측 점에 대해 계측을 하였다. 계측은 피부를 기준으로 계측을 하고, 폐에 대해서 영상을 이용하여 계측을 하였다. 영상으로 계측된 값을 통해 보상체 두께를 결정하였다. 또한 영상을 통한 보상체 위치를 결정하 고 확인하였다. 선량은 치료실에서 열형광소자를 이용하며 측정하였다. 결 과 : 신체 계측점에서 두께에 대하며 $1{\sim}2cm$의 일반 계측과 영상 계측의 차이를 보였다. 신체 길이는 일반 계측과 영상 계측이 $3{\sim}4cm$의 차이가 있었다. 또한, 영상을 통해 보상체의 밑그림을 그릴 수 있었다. 열형광소자를 이용한 선량 측정 결과 값은 머리, 목, 액와 ,가슴(폐 포함), 무릎 부위는 $92{\sim}98%$로 측정되었고 배, 골반, 서혜부, 발 부위는 $102{\sim}109%$로 측정되었다. 결 론 : TBI의 자세 잡이 과정에서 전산화 단층 모의치료기의 영상을 이용하는 것은 유용하였다. 신체 계측 점두께는 일반 계측과 영상계측의 차이가 적게 계측되었고, 길이에 대한 계측은 정확하게 이루어졌다. 영상을 이용한 계측은 신체 밀도까지 고려한 다양한 보상체 제작이 가능성을 보여주었다. 또한 보상체 위치 잡이를 정확하게 할 수 있고, 보상체의 모 양도 제한 받지 않고 쉽게 제작할 수 있다. 전체적인 치료실 자세 잡이 시간을 $15{\sim}20$분 단축할 수 있었고, 보상체 제작 시간은 약 15분 감소시켰다.

Keywords