Contrast-Detail Phantom을 이용한 CR에서 Image Plate의 사용 횟수에 따른 Contrast-Detail Curve의 변화

  • 이승철 (아산재단 서울아산병원 방사선팀) ;
  • 박장흠 (아산재단 서울아산병원 방사선팀) ;
  • 김재동 (아산재단 서울아산병원 방사선팀) ;
  • 박창현 (아산재단 서울아산병원 방사선팀)
  • 발행 : 2005.12.01

초록

Purpose : Image plate (IP) is substituted for film in computed radiography. This study is to investigate into a variation of contrast and detail by the number used of image plate in computed radiography. Materials and Methods : A Contrast-Detail(CD)-RAD 2.0 phantom(Nijmegen hospital, The Netherlands) was used for this study. The computed radiography(CR) CD-RAD phantom images were acquired at 40 kVp, 160 mA, 1.6 mAs, and small focus with the Shimadzu general radiography UD-150B-10 system and Fuji FCR 5000 image process system with speed of 200. The IP used including once, 5000 times, and 10000 times also was used. The numerical value of image quality figures (IQF) was produced by CD-RAD analyser(the program is installed in the directory), and then contrast-detail curve was drawn. Results : In this study, the value of IQF was 3.53 in IP used once, 3.40 in 5000 times, and 3.22 in 10000 times. Conclusions : There was a variation of contrast-detail curve by the number used of IP with contrast-detail phantom in computed radiography. Therefore, it is necessary that the IP with lower IQF and a shift of contrast-detail curve to the lower left part is used.