임플랜트 지대주의 삭제과정이 결합부 안정성에 미치는 영향

THE EFFECT OF PREPARATION PROCEDURE ON IMPLANT-ABUTMENT JOINT STABILITY

  • 이장욱 (서울대학교 치과대학 치과보철학교실) ;
  • 김창회 (서울대학교 치과대학 치과보철학교실) ;
  • 장경수 (서울대학교 치과대학 치과보철학교실) ;
  • 임영준 (서울대학교 치과대학 치과보철학교실)
  • Lee Jang-Wook (Department of Prosthodontics, Graduate School, Seoul National University) ;
  • Kim Chang-Whe (Department of Prosthodontics, Graduate School, Seoul National University) ;
  • Jang Kyung-Soo (Department of Prosthodontics, Graduate School, Seoul National University) ;
  • Lim Young-Jun (Department of Prosthodontics, Graduate School, Seoul National University)
  • 발행 : 2005.10.01

초록

Statement of problem: Little is known about the effect of abutment preparation procedure on do-torque values in different implant platform and the relationship of final do-torque values with different implant platform size. Purpose: This study evaluated the effect of abutment preparation procedure on do-torque values in different implant platform and the relationship of final do-torque values with different implant platform size. Material and method: Six ITI implants (2 narrow-neck implants, 2 regular-neck implants, 2 wide-neck implants) and six Branemark implants (2 narrow platforms, 2 regular platforms, 2 wide platforms) were embedded in each acrylic resin block with epoxy resin. Eighteen $synOcta^(R)$ abutments (6 narrow-neck implant-abutments, 6 regular-neck implant-abutments, 6 wide-neck implant-abutments) and eighteen esthetic abutments (6 narrow platform-abutments, 6 regular platform-abutments, 6 wide platform-abutments) were tightened to each implant with digital torque gauge. Initial do-torque values were measured using digital torque gauge. After preparation of abutments, Final do-torque values were measured with digital torque gauge. Results and conclusion: 1. Screws loosening or abutments motion were not detected in all experimental group, but some scratches of implant-abutment joints were detected in all group 2. Reduction ratios of final do-torque values were greater than initial do-torque values in all measured group, except in narrow-neck implant-abutment group (p<0.05). 3. Reduction ratios of final do-torque values in wide-neck implant-abutment group were greater than regular-neck implant-abutment group (p<0.01). 4. The greatest standard deviation value was detected in wide platform group in both implant systems.

키워드

참고문헌

  1. Branemark PI. Osseointegration and its experimental background. J Prosthet Dent 1983;50:399-410 https://doi.org/10.1016/S0022-3913(83)80101-2
  2. Zarb GA, Schmitt A. The longitudinal clinical effectiveness of osseointegrated dental implants: the Toronto study. Part III: Problems and complications encountered. J Prosthet Dent 1990;64:185-94 https://doi.org/10.1016/0022-3913(90)90177-E
  3. Jemt T, Laney WR, Harris D, Henry PJ, Krogh PH Jr, Polizzi G, et al. Osseointegrated implants for single tooth replacement: a 1-year report from a multicenter prospective study. Int J Oral Maxillofac Implants 1991;6:29-36
  4. Binon PP. The effect of implant/abutment hexagonal misfit on screw joint stability. Int J Prosthodont 1996;9:149-60
  5. Jorneus L, Jemt T, Carlsson L. Loads and designs of screw joints for single crowns supported by osseointegrated implants. Int J Oral Maxillofac Implants 1992;7:353-9
  6. Zhang F, Zhao Y, Wang H, Razzoog ME, Lang BR. The preparation procedure of TiAdapt abutment effect on the implant to abutment joint stability. Hua Xi Yi Ke Da Xue Xue Bao 2002;33:247-9
  7. Martin WC, Woody RD, Miller BH, Miller AW. Implant abutment screw rotations and preloads for four different screw materials and surfaces. J Prosthet Dent 2001;86:24-32 https://doi.org/10.1067/mpr.2001.116230
  8. Hagiwara M, Chashi N. New tightening technique for threaded fasteners. Materials and Engineering Proceedings of the International Offshore Mechanics and Artic Engineering Symposium, vol 3. New York: American Society of Mechanical Engineers 1992;371-76
  9. Tzenakis GK, Nagy WW, Fournelle RA, Dhuru VB. The effect of repeated torque and salivary contamination on the preload of slotted gold implant-prosthetic screws. J Prosthet Dent 2002;88:183-91 https://doi.org/10.1067/mpr.2002.127604
  10. Haack JE, Sakaguchi RL, Sun T, Coffey JP. Elongation and preload stress in dental implant abutment screws. Int J Oral Maxillofac Implants 1995;10:529-36
  11. Schulte JK, Coffey J. Comparison of screw retention of nine abutment systems: a pilot study. Implant Dent 1997;6:28-31
  12. Khraisat A, Stegaroiu R, Nomura S, Miyakawa O. Fatigue resistance of two implant/abutment joint designs. J Prosthet Dent 2002;88:604-10 https://doi.org/10.1067/mpr.2002.129384
  13. Norton MR. Assessment of cold welding properties of the internal conical interface of two commercially available implant systems. J Prosthet Dent 1999;81:159-66 https://doi.org/10.1016/S0022-3913(99)70243-X
  14. Cantwell A, Hobkirk JA. Preload loss in gold prosthesis-retaining screws as a function of time. Int J Oral Maxillofac Implants 2004;19:124-32
  15. Al Rafee MA, Nagy WW, Fournelle RA, Dhuru VB, Tzenakis GK, Pechous CE. The effect of repeated torque on the ultimate tensile strength of slotted gold prosthetic screws. J Prosthet Dent 2002;88:176-82 https://doi.org/10.1067/mpr.2002.127652
  16. Westland IA. The energy requirements of the dental cutting process. J Oral Rehabil 1980;7:51-63 https://doi.org/10.1111/j.1365-2842.1980.tb01463.x
  17. Blue DS, Griggs JA, Woody RD, Miller BH. Effects of bur abrasive particle size and abutment composition on preparation of ceramic implant abutments. J Prosthet Dent 2003;90:247-54 https://doi.org/10.1016/S0022-3913(03)00368-8
  18. Lee J, Kim YS, Kim CW, Han JS. Wave analysis of implant screw loosening using an air cylindrical cyclic loading device. J Prosthet Dent 2002;88:402-8 https://doi.org/10.1067/mpr.2002.128099
  19. Hoyer SA, Stanford CM, Buranadham S, Fridrich T, Wagner J, Gratton D. Dynamic fatigue properties of the dental implant-abutment interface: joint opening in wide-diameter versus standard-diameter hex-type implants. J Prosthet Dent 2001;85:599-607 https://doi.org/10.1067/mpr.2001.115250
  20. Gratton DG, Aquilino SA, Stanford CM. Micromotion and dynamic fatigue properties of the dental implant-abutment interface. J Prosthet Dent 2001;85:47-52 https://doi.org/10.1067/mpr.2001.112796
  21. Merz BR, Hunenbart S, Belser UC. Mechanics of the implant-abutment connection: an 8-degree taper compared to a butt joint connection. Int J Oral Maxillofac Implants 2000;15:519-26
  22. Jansen VK, Conrads G, Richter EJ. Microbial leakage and marginal fit of the implant-abutment interface. Int J Oral Maxillofac Implants. 1997 Jul-Aug;12(4): 527-40. Erratum in: Int J Oral Maxillofac Implants 1997;12:709