A Synthetic Aperture Focusing Method for Three-way Dynamic Focusing

  • Published : 2005.02.01

Abstract

A novel synthetic aperture method for real-time three-way dynamic focusing is proposed, which provides lateral beam patterns represented as the product of Fourier transforms of transmit subaperture, receive subaperture, and a synthetic window function. In the proposed method, all array elements are fired individually and for each firing echo signals are recorded from all elements of a receive subaperture moving along an array with the transmit element. The three-way dynamic focusing is then achieved by employing a synthetic aperture algorithm for two-way dynamic focusing and a synthetic focusing method for transmit dynamic focusing. Both theoretical analysis and computer simulation results show that the proposed method produces ultrasound beams with improved lateral resolution at all depths compared to the conventional phased array imaging and synthetic aperture focusing methods.

Keywords

References

  1. Levin F. Nock, and Gregg E. Trahey, 'Synthetic Receive Aperture Imaging with Phase Correction for Motion and for Tissue Inhomogeneities-Part ': Basic Principles', IEEE Trans. Ultrason., Ferroelec., Freq. Contr., Vol. 39, No.4, pp. 489-495, 1992 https://doi.org/10.1109/58.148539
  2. Catherine H. Frazier, and William D. O'Brien, 'Synthetic Aperture Techniques with a Virtual Source Element', IEEE Trans. Ultrason., Ferroelec., Freq. Contr., Vol. 45, No.1, pp. 196-207, 1998 https://doi.org/10.1109/58.646925
  3. Mustafa Karaman, and Matthew O'Donnell, 'Synthetic Aperture Imaging for Small Scale Systems', IEEE Trans. Ultrason., Ferroelec., Freq. Contr., Vol. 42, No.3, pp. 429-442, 1995 https://doi.org/10.1109/58.384453
  4. Mustafa Karaman, and Matthew O'Donnell, 'Subaperture Processing for Ultrasonic Imaging', IEEE Trans. Ultrason., Ferroelec., Freq. Contr., Vol. 45, No.1, pp. 126-135, 1998 https://doi.org/10.1109/58.646917
  5. Richard Y. Chiao, Lewis J. Thomas, and Seth D. Silverstein, 'Sparse Array Imaging with Spatially-Encoded Transmits', Proc. IEEE Ultrason. Symp., pp. 1679-1682, 1997
  6. Moo-Ho Bae, and Mok-Kun Jeong, 'A Study of Synthetic-Aperture Imaging with Virtual Source Elements in B-Mode Ultrasound Imaging Systems', IEEE Trans. Ultrason., Ferroelec., Freq. Contr., Vol. 47, No.6, pp. 1510-1519, 2000 https://doi.org/10.1109/58.883540
  7. Geoffrey R. Lockwood, James R. Talman, and Shelby S. Brunke, 'Real-Time 3-D Ultrasound Imaging Using Sparse Synthetic Aperture Beamforming', IEEE Trans. Ultrason., Ferroelec., Freq. Contr., Vol. 45, No.4, pp. 980-988, 1998 https://doi.org/10.1109/58.710573
  8. S. W. Flax, and M. O'Donnell, 'Phase-Aberration Correction Using Signals From Point Reflectors and Diffuse Scatterers: Basic Principles', IEEE Trans. Ultrason., Ferroelec., Freq. Contr., Vol. 35, No.6, pp. 758-767, 1988 https://doi.org/10.1109/58.9333
  9. Gregg E. Trahey, and Levin F. Nock, 'Synthetic Receive Aperture Imaging with Phase Correction for Motion and for Tissue Inhomogenities-Part ' .: Effects of and Correction for Motion' IEEE Trans. Ultrason., Ferroelec., Freq. Contr., Vol. 39, No.4, pp. 496-501, 1992 https://doi.org/10.1109/58.148540
  10. B. D. Steinberg, Principals of Aperture and Array System Design. New York: J. Wiley, 1976