DOI QR코드

DOI QR Code

다수의 계수구역간의 검사체적을 이용한 소형 세포농도센서

Micro Cell Counter Using a Fixed Control Volume Between Double Electrical Sensing Zones

  • 이동우 (한국과학기술원 바이오시스템학과, 디지털나노구동연구단) ;
  • 이소연 (한국과학기술원 바이오시스템학과, 디지털나노구동연구단) ;
  • 조영호 (한국과학기술원 바이오시스템학과 및 기계공학과, 디지털나노구동연구단)
  • 발행 : 2005.12.01

초록

We present a novel flow-rate independent cell counter using a fixed control volume between double electrical sensing zones. The previous device based on the single electrical cell sensing in a given flow-rate requires an accurate fluid volume measurement or precision flow rate control. The present cell counter, however, offers the flow-rate independent method for the cell concentration measurement with counting cells in a fixed control volume of $22.9{\pm}0.98{\mu}{\ell}$. In the experimental study, using the RBC (Red Blood Cell), we have compared the measured RBC concentrations from the fabricated devices with those from Hemacytometer. The previous and present devices show the maximum errors of $20.3\%\;and\;16.1\%$, which are in the measurement error range of Hemacytometer (about $20\%$). The present device also shows the flow-rate independent performance at the constant flow-rates ($5{\mu}{\ell}/min$ and $10{\mu}{\ell}/min$) and the varying flow-rate (4, 2, and $4{\mu}{\ell}/min$). Therefore, we demonstrate that the present cell counter is a simple and automated method for the cell concentration measurement without requiring an accurate fluid measurement and precision flow-rate control.

키워드

참고문헌

  1. Koch, M., Evans, A.G.R. and Brunnschweiler, A., 1999, 'Design and Fabrication of a Micromachined Coulter Counter,' Journal of Micromechanics and Microengineering., Vol. 9, pp. 159-161 https://doi.org/10.1088/0960-1317/9/2/313
  2. Larsen, U.D., Blankenstein, G. and Branebjerg, J., 1997, 'Microchip Coulter Particle Counter,' Proc. Int. Conf. On Solid-State Sensors and Actuators (Transducers '97), pp. 1319-1322 https://doi.org/10.1109/SENSOR.1997.635479
  3. Son, S.U., Choi, Y.H. and Lee, S.S., 2003, 'Fabrication of Micro Cell Counter Integrated with Oxygen Micropump,' Proceedings of the Micro Total Analysis Systems 2003, Boston, USA, pp. 179-182
  4. Lee, G.-B., Fu, L.-M., Yange, R.-J. and Pan, Y.-J., 'Micro Flow Cytometers Using Electrokinetic Forces with Integrated Optical Fibers for On-Line Cell/Particle Counting and Sorting,' Proceedings of the Micro Total Analysis Systems 2003, Boston, USA, pp. 45-48
  5. Allen, T., 1990, Particles Size Measurements, Chapman and Hall, London, New York
  6. Prescott, L.M., Harley, J.P. and Klein, D.A., 2002, Microbiology, McGraw-Hill, New York