Quantitative Assay of Hepatitis B Surface Antigen by Using Surface Plasmon Resonance Biosensor

  • Published : 2005.08.01

Abstract

We performed a basic experiment for the rapid, on-line, real-time measurement of hepatitis B surface antigen using a surface plasmon resonance biosensor. We immobilized anti­HBsAg (hepatitis B surface antigen) polyclonal antibody, as a ligand, to the dextran layer on a CM5 chip surface that had previously been activated by N-hydroxysuccinimide. A sample solution containing HBsAg was fed through a microfluidic channel, and the reflecting angle change due to the mass increase from the binding was detected. The binding characteristics between HBsAg and its polyclonal antibody followed the typical monolayer adsorption isotherm. When the entire immobilized antibody had interacted, no additional, non-specific binding occurred, suggesting the immunoreaction was very specific. The bound antigen per unit mass of the antibody was independent of the immobilized ligand density. No significant steric hindrance was observed at an immobilization density of approximately $17.6 ng/mm^2$. The relationship between the HBsAg concentration in the sample solution and the antigen bound to the ligand was linear up to ca. $40{\mu}g$/mL. This linearity was much higher than that of the ELISA method. It appeared the anti­gen-antibody binding increased as the immobilized ligand density increased. In summary, this study showed the potential of this SPR biosensor-based method as a rapid, simple and multi­sample on-line assay. Once properly validated, it may serve as a more efficient method for HBsAg quantification for replacing the ELISA.

Keywords

References

  1. Rich, R. L. and D. G. Myszka (2000) Advances in surface plasmon resonance biosensor analysis. Curr. Opin. Biotechnol. 11: 54-61 https://doi.org/10.1016/S0958-1669(99)00054-3
  2. Myszka, D. G. and R. L. Rich (2000) Implementing surface plasmon resonance biosensors in drug discovery. PSTT 3: 310-316 https://doi.org/10.1016/S1461-5347(00)00288-1
  3. Bianchi, N., C. Rutigliano, M. Tomassetti, G. Feriotto, F. Zorzato, and R. Gambari (1997) Biosensor technology and surface plasmon resonance for real-time detection of HIV-1 genomic sequences amplified by polymerase chain reaction. J. Clin. Virol. 8: 199-208 https://doi.org/10.1016/S0928-0197(97)00025-1
  4. Baek, S. H., Y. B. Shin, M. G. Kim, H. S. Ro, E. K. Kim, and B. H. Chung (2004) Surface plasmon resonance imaging analysis of hexahistidine-tagged protein on the gold thin film coated with a calix crown derivative. Biotechnol. Bioprocess Eng. 9: 143-146 https://doi.org/10.1007/BF02932998
  5. Nice, E. C., T. L. McInerney, and D. C. Jackson (1996) Analysis of the interaction between a synthetic peptide of influenza virus hemagglutinin and monoclonal antibodies using an optical biosensor. Mol. Immunol. 33: 659-670 https://doi.org/10.1016/0161-5890(96)00016-8
  6. Nelson, R. W., D. Nedelkov, and K. A. Tubbs (2000) BIACORE and mass spectrometry: Identification of epitope-tagged proteins in E. coli lysates. BIAjournal 7: 25-26
  7. Tung, J. S., J. Gimenez, C. T. Przysiecki, and G. Mark (1998) Characterization of recombinant hepatitis B surface antigen using surface plasmon resonance. J. Pharm. Sci.-US 7: 76-80 https://doi.org/10.1021/js970154l
  8. Oh, B. K., W. C. Lee, W. H. Lee, and J. W. Choi (2003) Nano-scale probe fabrication using self-assembly technique and application to detect of Escherichia coli O157:H7. Biotechnol. Bioprocess Eng. 8: 227-232 https://doi.org/10.1007/BF02942270
  9. Gomara, M. J., G. Ercilla, M. A. Alsina, and I. Haro (2000) Assessment of synthetic peptides for hepatitis A diagnosis using biosensor technology. J. Immunol. Methods 246: 13-24 https://doi.org/10.1016/S0022-1759(00)00295-7
  10. Bruno, J. (1998) Validating BIACORE assays. BIAjournal 5: 9-11
  11. Magnusson, R., B. Rosenbrand, M. Wouda, and E. Mascher (1997) Quantitative process analysis of pantarin using BIACORE probe. BIAjournal Special Issue: 31
  12. Karlsson, R., H. Roos, J. Bruno, and L. Stolz (1997) Practical aspects concerning direct detection of low molecular weight analytes using BIACORE 2000. BIAjournal Special Issue: 18-21
  13. Bazarsuren, A. and S. Panzner (1999) Re-use of thiol coupled surfaces at different ligand concentrations. BIAjournal 6: 25-26
  14. Haemaelaeinen, M. and F. Markey (1998) Sorting the sheep from goats: Screening for low molecular weight binders with BIACORE. BIAjournal 5: 12-17
  15. Takacs, M. A., S. J. Jacobs, R. M. Bordens, and S. J. Swanson (1999) Detection and characterization of antibodies to PEG-IFN-$\alpha$2b using surface plasmon resonance. J. Interf. Cytok. Res. 19: 781-789 https://doi.org/10.1089/107999099313631