Changes in Pressure-Flow Control Characteristics of Shunt Valves by Intracranial Pressure Pulsation: an In Vitro Study

  • Lee, Chong-Sun (College of Mechanical and Control Engineering, Handong Global University) ;
  • Kim, Joo-Young (College of Mechanical and Control Engineering, Handong Global University)
  • Published : 2005.08.01

Abstract

Shunt valves used to treat patients with hydrocephalus were tested to investigate influence of intracranial pressure pulsation on their flow control characteristics. Five commercial shunt valves were tested in the flow loop that simulates pulsed flow under pressure pulsation. As 20cc/hr of flow rate was adjusted at a constant pressure, application of $40mmH_2O$ of pressure pulse increased the flow rate by $67.9\%.$ As a 90cm length catheter was connected to the valve outlet, increase in the flow rate was substantially reduced to $17.5\%.$ As the flow rate was adjusted to 40cc/hr at a constant pressure, increase in the flow rate was $51.1\%$ with the same pressure pulsation of $40mmH_2O$. The results indicated that pressure-flow control characteristics of shunt valves implanted above human brain ventricle is quite different from those obtained by syringe pump test at constant pressures right after manufacture. The influence of pressure pulsation was observed to be more significant at low flow rate and the flexibility of the outlet silicone catheter was estimated to significantly reduce flow increase due to pressure pulsation.

Keywords

References

  1. Drake JM, Saint-Rose C, The Shunt Book, New York Blackwell Scientific, 1995
  2. Post, E. M, 'Currently available shunt systems: a review', Neurosurgery, Vol. 16, pp. 257-260, 1985 https://doi.org/10.1227/00006123-198502000-00027
  3. Neurological implants: streile, single use hydrocephalus shunts and components, ISO7197, 2nd. ed., 1997
  4. Lee, C.S., Kim, Y.H., Jang, J.Y., Suh, C.M., Yang, K.S., Sohn, C.H., Ra, Y.S., Kim, Y.S., 'Development of shunt valves used to treat patients with hydrocephalus', Journal of Biomedical Engineering Research, Vol. 22, pp. 431-438, 2001
  5. Hong, I.S., Lee, C.S., Jang, J.Y., 'Changes in the pressure-flow control characteristics of shunt valves by intracranial pressure pulsation', Journal of Biomedical Engineering Research, Vol.23, pp. 391-395, 2002
  6. Cardoso ER, Rowan JO, 'Analysis of the cerebrospinal fluid pulse wave in intracranial pressure', J. Neurosurg., Vol. 59, pp. 817-821, 1983 https://doi.org/10.3171/jns.1983.59.5.0817
  7. Ursino, M., Iezzi, M., Stocchetti, N., 'Intracranial pressure dynamics in patients with acute brain damage', IEEE Biomedical Engineering, Vol.42, pp.529-540, 1995 https://doi.org/10.1109/10.387192
  8. Jang, J.Y., Lee, C.S., Suh, C.M., 'Numerical simulation of the pressure-flow control characteristics of shunt valves used to treat patients with hydrocephalus', Journal of Biomedical Engineering Research, Vol. 23, pp. 403-411, 2001
  9. Won, C.S., Huh, N.K, Lee, C.S., 'Flow structure analysis for designing medical CSF valves' , Korean Computational Fluid Mechanics, Vol. 6, pp. 40-46, 2001