Neutron Generation from a 24 MV Medical Linac

24 MV 의료용 선형가속기의 중성자 발생에 관한 연구

  • Jeong Dong Hyeok (Department of Radiation Oncology, Wonkwang University Hospital) ;
  • Kang Jeong Ku (Department of Radiation Oncology, Presbyterian Medical Center, Jeonju, Korea) ;
  • Lee Jeong Ok (Department of Radiotechnology, Wonkwang Health Science College)
  • 정동혁 (원광대학병원 방사선종양학과) ;
  • 강정구 (전주예수병원 방사선종양학과) ;
  • 이정옥 (원광보건대학 방사선과)
  • Published : 2005.06.01

Abstract

The energy spectra and dose calculations were performed for secondary neutrons from a 24 MV LINAC using MCNPX code (V2, 4, 0). The energy spectra for neutrons and photons emitted from the LINAC head, and absorbed dose to water were calculated in water phantom. The absorbed doses calculated with Monte Carlo were $0.66\~0.35$ mGy/photon Gy at the surface to d=5 cm, and calculated with interaction data was 0.52 mGy/photon Gy at the depth of electron equilibrium in water. We have shown that this work can be applied to dose estimation of neutrons from high energy LINAC through the comparison of our results with other results.

본 연구에서는 몬테칼로 계산(MCNPX 2.4.0)을 이용하여 24MV 의료용 선형가속기에서 발생하는 이차중성자의 에너지분포와 선량계산을 수행하였다. 먼저 선형가속기 두부(head)에서 방출되는 광자와 중성자의 에너지분포를 계산하였으며, 이를 이용하여 물팬텀에서 광자와 중성자의 흡수선량을 계산하였다. 몬테칼로 계산결과 물표면에서 5cm까지 중성자의 흡수선량은 $0.66\~0.35mGy/photon\;Gy$로 나타났으며, 상호작용 자료를 이용한 근사적 계산결과는 전자평형 깊이에서 0.52mGy/photon Gy로 나타났다. 다른 연구결과들과 비교를 통하여 본 연구가 고에너지 선형가속기에서 발생되는 중성 자의 선량평가에 응용될 수 있음을 제시하였다.

Keywords