DOI QR코드

DOI QR Code

Development of an automatic measurement system for the AC-DC current transfer difference of the thermal current converter

열전형 전류 변환기의 교류-직류 전류 변환차이 자동측정시스템 개발

  • Published : 2005.09.30

Abstract

We have developed a dual-channel type automatic measurement system to evaluate AC-DC current transfer difference of the thermal current converter(TCC) which is primary standard of AC current. The output drift effect of the TCC is minimized by measuring simultaneously the output voltages of two TCCs using voltmeter. Furthermore, the offset voltage of the voltmeter is cancelled nearly out by taking the average values of two outputs of TCCs measured with the forward-reverse directions using dual channel scanner. The uncertainties of the automatic system were 7 to $86{\mu}A/A$ for 3 mA to 10 A at 40 Hz to 20 kHz, which were evaluated by the comparisons between adjacent range of TCCs and inter-comparison with national measurement institute of Germany(PTB). The capability for ac-dc transfer difference measurement was improved by one order compared with that for the manual ac-dc measurement system.

Keywords

References

  1. F. L. Hermach, 'Thermal converters as AC-DC transfer standards for current and voltage measurements at audio frequencies', J Res. of NBS, vol. 48, no. 2, pp. 121-138, 1952
  2. E. S. Williams, 'Thermal current converters for accurate AC current measurement', IEEE Trans. Instrum. Meas.. vol. IM-25, no. 4, pp. 519-523, 1976 https://doi.org/10.1109/TIM.1976.6312279
  3. B. D. Inglish, 'AC-DC transfer standards-present status and future directions', IEEE Trans. Instrum. Meas. vol. IM-34, no. 2, pp. 285-290, 1985 https://doi.org/10.1109/TIM.1985.4315326
  4. B. D. Inglish, 'Standards for AC-DC transfer', Metrologia, vol. 29, pp. 191-199, 1992 https://doi.org/10.1088/0026-1394/29/2/007
  5. M. Klonz, 'Current developments in accurate AC-DC transfer measurements', IEEE Trans. Instrum. Meas., vol. IM-44, no. 2, pp. 363-366, 1995
  6. J. R. Kinard, T. E. Lipe, and C. B. Childers, 'Extension of the NIST AC-DC difference calibration service for current to 100 kHz', J. Res. Natl. Inst. Stand. Technol., vol. 102, pp. 75-83, 1997 https://doi.org/10.6028/jres.102.007
  7. K. E. Rydler and V. Tarasso, 'Current AC-DC transfer measurements in the microampere range', IEEE Trans. Instrum. Meas. vol. IM-50, no. 2, pp. 342-344, 2001 https://doi.org/10.1109/19.918137
  8. E. S. Williams and J. Kinard, 'A dual-channel automated comparator for AC-DC difference measurements', IEEE Trans. Instrum. Measu., vol. IM-34, no. 2, pp. 290-294, 1985 https://doi.org/10.1109/TIM.1985.4315327
  9. K. R. Rydler, 'High precision automated measuring system for AC-DC current transfer standards', IEEE Trans. Instrum. Measu.. vol. IM-42, no. 2, pp. 608-611, 1993 https://doi.org/10.1109/19.278637
  10. I. Budovsky, 'AC-DC difference calibrations on resistors at NML', Proceedings, Metrology Society of Australia Biennial Conference, Melbourne, pp. 129-133, 1997
  11. 권성원, 이래덕,M. Klonz, '교류전류 l차표준용 열 전형 전류변환기의 제작', 센서학회지, 제1권, 제1호, pp 77-83, 1992
  12. J. R. Kinard, T. E. Lipe, and C. B. Childers, 'AC-DC difference relationships for current shunt and thermal converter combinations', IEEE Trans. Instrum. Measu.. vol. IM-40, no. 2, pp. 352-355, 1991 https://doi.org/10.1109/TIM.1990.1032958
  13. 권성원, 조용명, 김규태, 강전홍, 박영태, '교류-직류 변환오차 자동 측정시스템' 센서학회지, 제7권, 제 6호, pp. 401-408, 1998
  14. D. N. Homan, 'Applications of coaxial chokes to AC bridge circuits', J. Res. of NBS, vol. 72c, pp. 161, 1968
  15. ISO/IEC/OIML/BIPM, 'Guide to the Expression of Uncertainty in Measurement', 1993(E)
  16. M. Klonz, et al., 'AC-DC curent transfer step-up and step-down calibration and uncertainty calculation', IEEE Trans. Instrum. Measu.. vol. IM-51, no. 5, pp. 1027-1034, 2002 https://doi.org/10.1109/TIM.2002.806008