A Study on the exposure dose for the computed tomography

컴퓨터 단층촬영시 환자피폭선량에 관한 연구

  • Published : 2004.06.30

Abstract

This study was conducted to estimate absorbed radiation doses associated with CT examinations. We compared CT dose index between single detector CT and multi detector CT. To establish radiation dose criteria in CT examination in Korea, we measured radiation dose for CT examinations in Seoul and kyungki-do. The results obtained were as follows ; 1. Averaged CTDIW value per 100 mAs was $13.5{\pm}3.2\;mGy$, and ranged from 8.1 mGy to 19.1 mGy in head phantom, was $7.1{\pm}2.0\;mGy$, and ranged from 3.7 mGy to 10.9 mGy in body phantom. 2. CTDIW was 3.2 mGy(1.26 times) larger in multi detector CT than single detector CT in head phantom, and 2.1 mGy(1.34 times) larger in body phantom. 3. The dose was the highest in 4 channel multi detector CT, and followed 8 channel multi detector CT, 16 channel multi detector CT and single detector CT in head phantom. And the dose was the highest in 4 channel and 8 channel multi detector CT, and followed 16 channel multi detector CT and single detector CT in body phantom.

본 연구에서는 현재 국내에서 사용되고 있는 여러 기종의 CT장치를 대상으로 하여 CT검사로 인한 방사선피폭 정도를 실험을 통하여 알아보고, 외국의 사례와 비교함으로써 CT장치의 성능관리의 하나인 피폭선량 기준 설정에 필요한 기초 데이터를 제시하고자 서울시 및 경기도에 위치한 병의원 및 종합병원에서 가동 중인 32대의 CT장치를 대상으로 CTDI값을 측정한 결과 다음과 같았다. 1) Head phantom의 100 mAs 당 $CTDI_W$값은 $8.1{\sim}19.1\;mGy$ 범위였고, 평균 $13.5{\pm}3.2\;mGy$였다. 그리고 body phantom의 $CTDI_W$값은 $3.7{\sim}10.9\;mGy$ 범위였고, 평균 $7.1{\pm}2.0\;mGy$였다. 2) Single detector CT와 multi detector CT의 $CTDI_W$값을 비교해 보면, multi detector CT가 single detector CT에 비해 head phantom에서는 평균 3.2 mGy(약 1.26배), body phantom에서는 평균 2.1 mGy(약 1.34배) 높았다. 3) Channel 수에 따른 $CTDI_W$값 비교에서는 head pahantom에서는 4 channel CT가 가장 높았으며, 8 channel CT, 16 channel CT, single detector CT순이었으며, body phantom에서는 역시 4 channel CT와 8 channel CT, 16 channel CT, single detector CT순이었다.

Keywords