저잡음 심자도측정시스템 개발을 위한 고온초전도 SQUID 센서의 제작

Fabrication of HTS SQUID Sensors for the Application to a High S/N Ratio Magnetocardiograph System

  • Kim I. S (Korea Research Institute of Standards and Science,) ;
  • Yu K. K (Korea Research Institute of Standards and Science,) ;
  • Park Y. K (Korea Research Institute of Standards and Science,)
  • 발행 : 2004.10.01

초록

YBCO do superconducting quantum interference device (SQUID) magnetometers based on bicrystal Josephson junctions on 10 mm ${\times}$ 10 mm $SrTiO_3$ substrates have been fabricated. The pickup coil of the device was designed to have 16 parallel loops with 50-fm-wide lines. We could obtain optimised direct coupled YBCO SQUID magnetometer design with field sensitivity $B_{N}$ $\Phi$/ of $4.5 nT/\Phi_{0}$ and magnetic field noise $B_{N}$ of about $22 fT/Hz^{1}$2/ with an I/f corner frequency of 2 Hz measured inside a magnetically shielded room. Preliminary results of magnetocardiograph measurement using the HTS SQUID magnetometers show signal to noise ratio of about 110, which is comparable to the quality of a commercial MCG system based on Nb-SQUIDs.

키워드