A new ion beam extraction system is designed using a simple ion mass filter and a micro mass balance and a QMS based detecting system. A quadrupole Mass Filter is used for selective ion beam formation from inductively coupled high density plasma sources with appropriate electrostatic lens and final analyzing QMS. Also a quartz crystal microbalance is set between a QMF and a QMS to measure the etching and polymerization rate of the mass selected ion beam. An inductively coupled plasma was used as a ion/radical source which had an electron temperature of 4-8 eV and electron density of $4${\times}$10^{11}$#/㎤. A computer interfaced system through 12bit AD-DA board can control the pass ion mass of the qmf by setting RF/DC voltage ratio applied to the quadrupoles so that time modulation of pass ion's mass is possible. So the direct measurements of ion - surface chemistry can be possible in a resolution of $1\AA$/sec based on the qcm's sensitivity. A full set of driving software and hardware setting is successfully carried out to get fundamental plasma information of the ICP source and analysed $Ar^{+}$ beam was detected at the $2^{nd}$ QMS.

Keywords

References

  1. J. W. Coburn, H. F. Winters, J. Appl. Phys., 50 (1979) 3189-3196 https://doi.org/10.1063/1.326355
  2. H. F. Winters, J. W. Coburn, Surface Science Reports, 14 (1992) 164-269
  3. H. F. Winters, J. W. Coburn, J. Vac. Sci. Tech., B9 (1991) 197-207
  4. David C. Gray, Vivek Mohindra, H. H. Sawin, J. Vac. Sci. Tech., A12(2) 354
  5. H. Sugai, K. Nakamura, Y. Hikosaka, M. Nakamura, 4141^{st}American Vacuum Society National Symposium Proceedings, (1994) 226
  6. H. Y. Chang, 97 Plasma Forum Program by Korean Vacuum Society, Jeju National University (1997)
  7. Laframboise, University of Toronto, Institute for Aero space Studies Report, No.100 (1966)
  8. 김명수, 질량분석학, 대우학술총서*자연과학45, 민음사, (1987) 38-42