The Design of an Infrared Transcutaneous Control Unit for Totally Implantable Middle Ear System

완전 이식형 인공중이를 위한 체외 및 체내 제어시스템 구현

  • Published : 2004.09.01

Abstract

An infrared remote control-type transcutaneous control device using a $\mu$-processor is design for the totally implantable middle ear system. An infrared light transmission model for the tissue of skin was introduced and then a radiant intensity and the required current of the infrared light emitting diode(IR LED) driving circuit at transmission part were calculated for the external control device. And the transmission part generates IR signal by the system's own data protocol which prevents interferences from other infrared remote controls of the household appliances. The control part of the implanted device was designed to analyze functions of the received infrared(IR) signal that indicate the power ON/OFF and volume UP/DOWN. After the system is implemented, a data transmission experiments using 4 mm thickness of porcine skin were carried out. From the experiment, it was verified that the infrared control signal was transmitted to receiving module of the implemented system without any error.

체내에 이식되는 완전 이식형 인공중이 시스템의 제어를 위해 저전력 마이크로프로세서인 MSP430을 이용한 적외선 리모컨형 외부제어기를 설계하였고, 수신된 신호 해석을 위한 제어 알고리즘을 설계하여 전원 ON/OFF와 소리크기 UP/DOWM 동작을 수행하도록 하였다. 먼저 피부 투과시 발생하는 적외선 신호의 반사 및 투과모델을 이용하여 시스템에 필요한 적외선 LED 방사강도를 계산하고, 인반 적외선 리모컨과의 오작동을 방지하기 위해 고유의 데이터 프로토콜을 설정하였다. 이식되는 수신부의 적외선 수신모듈은 자체 소비전류를 줄이기 위해 마이크로프로세서의 내부 타이머에 의한 25% 듀티비를 갖는 출력신호를 전원으로 사용하였다. 4mm 두께의 돼지피부를 이용한 피부투과 전송실험을 실시하여 구현된 송수신 장치로 신호가 전달됨을 확인하였다.

Keywords

References

  1. J. I. Suzuki, H. Shono, K. Koga, and T. Akiyama, 'Early Studies and the History of Development of the Middle Ear Implant in Japan,' Advances in Audiology, vol. 4, pp. 1-14, 1998
  2. R. L. Goode, M. L. Rosenbaum, and A. J. Maniglia, 'The History and Development of the Implantable Hearing Aid,' The Otolaryngologic Clinics of NorthAmerica, vol. 28, no. 1, pp. 1-16, February 1995
  3. B. S. Song, T. Y. Jung, S. P. CHAE, M. N. Kim, and J. H. Cho, 'Proposal and Evaluation of Vibration Transducer with Minimal Magnetic Field Interference for Use in IME System by in-vitro Experiment,' IEICE Transaction on Electronics, vol. E85-C, no. 6, pp. 1374-1377, June 2002
  4. 임형규, 정의성, 김민규, 송병섭, 조진호, '완전 이식형 인공중이 시스템을 위한 무선 전력전달 장치의 설계,' 제 28회 대한의용생체공학회 춘계학술대회, 2003. 5
  5. Eric K. chan, Vrian sorg, Dmitry Protsenko, Vichael O'Neil, Vassoud Motamdi, and Ashley J. Welch, 'Effects of Compression on Soft Tissue Optical Properties,' IEEE Journal of Selected Topics in Quantum Electronics, vol. 2, no. 4, December, 1996 https://doi.org/10.1109/2944.577320
  6. 박선호 편저, IrDA 규격 해설과 적외선 데이터 통신 설계, 국제테크노정보연구소, 2000
  7. A. Ishimaru, 'Wave propagation and scattering in random media,' Academic Press, vol. 1, pp. 191-201, 1978
  8. M. J. C. Van Gemert, Steven L. Jacques, H. H. C. M. Sterenborg, and W. M. Star 'Skin Optics,' IEEE Transactions on Biomedical Engineering, vol. 36, no. 12, December 1989 https://doi.org/10.1109/10.42108
  9. A. J. Welch, 'The thermal response of laser irradiatied tissue,' IEEE Journal of Quantum Electronics QE-20, pp. 1471-1481, 1984 https://doi.org/10.1109/JQE.1984.1072339
  10. A. E. Cerussi, A. J. Berger, F. Bevilacqua, et al, 'Sources of Absorption and Scattering Contrast for Near-Infrared Optical Mammography,' Academic Radiology, vol. 8, no. 3, pp. 211-218, March 2001 https://doi.org/10.1016/S1076-6332(03)80529-9
  11. 박선호 편저, 적외선 공간통신과 원격 제어시스템, 국제테크노정보연구소, 2000