Abstract
The LAL (Limulus amebocyte lysate) test for the detection and quantification of endotoxin is based on the gelation reaction between endotoxin and LAL from a blood extract of Limulus polyphemus. The test is labor intensive, requiring dedicated personnel, a relatively long reaction time (approximately 1 h), relatively large volumes of samples and reagents and the detection of the end-point is rather subjective. To solve these problems, a miniaturized LOC (lab-on-a-chip) prototype, 62mm (L) ${\times}$ 18 mm (W), was fabricated using PDMS (polydimethylsiloxane) bonded to glass. Using this prototype, in which 2mm (W) ${\times}$ 44.3mm (L) ${\times}$ 100 $\mu\textrm{m}$ (D) microfluidic channel was constructed, turbidometric and chromogenic assay detection methods were compared, and the chromogenic method was found the most suitable for a small volume assay. In this assay, the kinetic-point method was more accurate than the end-point method. The PDMS chip thickness was found to be minimized to around 2 mm to allow sufficient light transmittance, which necessitated the use of a glass slide bonding for chip rigidity. Due to this miniaturization, the test time was reduced from 1 h to less than 10 min, and the sample volume could be reduced from 100 to ca. 4.4 ${\mu}$L. In summation, this study suggested that the LOC using the LAL test principle could be an alternative as a semi-automated and reliable method for the detection of endotoxin.