A Study of Prediction of Gas Transfer rate in Intravascular Lung Assist Device

혈관 내 폐 보조장치에서의 산소전달속도 예측에 관한 연구

  • 김기범 (전북대학교 공과대학 화학공학부, 공학연구원 공업기술연구센터) ;
  • 나도춘 (전북대학교 공과대학 화학공학부, 공학연구원 공업기술연구센터) ;
  • 김성종 (익산대학 환경공업화학과) ;
  • 정인수 (한국과학기술정보연구원) ;
  • 정경락 (전북대학교 공과대학 화학공학부, 공학연구원 공업기술연구센터) ;
  • 권대규 (전북대학교 공과대학 생체정보공학부)
  • Published : 2004.03.01

Abstract

The purpose of this paper was to find out the proper equation to predict the gas transfer rate for designing intravenous artificial lung assist device. The prepared hollow fiber modules were examined under various experimental conditions through experimental modeling before inserted the artificial lung assist d $\varepsilon$ vice into as venous. As a result, we can estimate the gas transfer as a function of the packing density. The gas transfer obtained from the experiment was similar to that from the equation, confirming the usefulness equation. Therefore, we can conclude the gas transfer of the intravenous artificial lung assist device as a function of the packing density, and this functions are very useful for predicting the gas transfer of the intravenous artificial lung assist device.

본 연구의 목적은 정맥 내 폐 보조장치 설계에 있어서 기체전달을 예측하기 위하여 혈관 내 폐 보조장치를 정맥 내에 삽입하기 전, 여러 가지 설계조건에서 실험적 모델링을 통하여 예측식을 찾고자 시도하였다. 실험결과 중공사 충진율의 함수에 따라 기체전달을 예측할 수 있었다. 실험에 의하여 얻어진 기체전달은 예측식에 의한 기체전달과 유사하여, 식에 대한 신뢰성을 얻을 수 있었다. 그러므로, 중공사의 충진율의 함수에 따라 기체전달을 예측할 수 있으며, 혈관 내 폐 보조장치의 기체전달을 예측하는데 매우 유용하였다.

Keywords

References

  1. 고분자과학과 기술 v.7 no.2 의료용 중공사막 김성수;김재진
  2. ASAIO J. v.45 Ex-vivo testing of the intravenous membrane oxygenator (IMO) W.J.Federspiel;L.W.Lund;J.A.Bultman;S.Wanant;J.Matoney;J.F.Golob;B.J.Frankowski;M.Watach;P.Litwak;B.G.Hattler
  3. ASAIO J. v.47 Development of an Implantable Artificial Lung: Challenges and Progress J.B.Zwischenberger;C.M.Anderson;K.E.Cook;S.D.Lick;L.F.Mockros;R.H.Bartlett https://doi.org/10.1097/00002480-200107000-00003
  4. 멤브레인 v.13 no.1 혈관 내 폐 보조장치설계를 위한 압력손실 특성 평가 김기범;권대규;박재관;정경락;이삼철
  5. Trans. Am. Soc. Artif. Intern. Organs v.33 An intravenacaval blood gas exchange (IVCBGE) device: a preliminary report J.D.Mortensen
  6. Artif. Organs v.16 Intravascular oxygenator: a new alternative method for augmenting blood gas transfer in patients with acute respiratory failure J.D.Mortensen https://doi.org/10.1111/j.1525-1594.1992.tb00271.x
  7. Annals of Biomedical Engineering v.26 Enhancement of O₂ and Co₂ transfer through microporous hollow fiber by pressure cycling V.Nodelman;H.Baskaran;J.S.;Ultman https://doi.org/10.1114/1.35
  8. ASAIO J. v.45 Ex-vivo testing of the intravenous membrane oxygenator (IMO) W.J.Federspiel;L.W.Lund;J.A.Bultman;S.Wanant;J.Matoney;J.F.Golob;B.J.Frankowski;M.Watach;P.Litwak;B.G.Hattler
  9. ASAIO J. v.40 Design and evaluation of a new low pressure loss implantable artificial lung S.N.Vaslef;K.F.Cook;R.J.:eonard;L.F.Mockros;R.W.Anderson https://doi.org/10.1097/00002480-199407000-00055
  10. Artificial Organs v.18 no.11 Computer-assisted design of an implantable intrathoracic artificial lung S.N.Vaslef;L.F.Mockros;K.E.Cook;R.J.Leonard;J.C.Sung;R.W.Anderson https://doi.org/10.1111/j.1525-1594.1994.tb03328.x
  11. Unit Operation of Chemical Engineering(5th ed.) W.L.McCabe;J.C.Smith;P.Harriott
  12. Transport Phenomena (2nd ed.) W.L.Beek;K.M.K.Muttzell;van J. W. Heuven
  13. Biotechnol. Prog. v.18 Mass Transfer in Blood Oxygenators Using Blood Analogue Fluids S.R.Wickramasinghe;C.M.Kahr;B.Han https://doi.org/10.1021/bp010192h
  14. Chem. Eng. J. v.3 Rate of blood oxygenation in flat plate membrane oxygenator S.Katoh;F.Yoshida
  15. Handbook of water analysis: Chemical parameter's L.M.L,Nollet;de Medina, H. L.
  16. Ph. D. dissertation Analysis and Designs of an Intravascular Lung Asisst Device S.N.Vaslef
  17. 유체역학(12판) 이택수
  18. Ph. D. dissertation, Northwestern University Oxygen diffusion in human blood plasma V.T.Ciuryla