The Journal of Korean Society for Radiation Therapy (대한방사선치료학회지)
- Volume 16 Issue 2
- /
- Pages.33-41
- /
- 2004
- /
- 1598-8449(pISSN)
A Comparison of coincidence between the Light field & the Radiation field using film and BIS
필름과 BIS 영상장치를 이용한 광/방사선조사야 일치성 비교평가
- Bang, Dong-Wan (Dept. of Radiation Oncology, Seoul National University Bundang Hospital) ;
- Seok, Jin-Yong (Dept. of Radiation Oncology, Seoul National University Bundang Hospital) ;
- Jeong, Yun-Ju (Dept. of Radiation Oncology, Seoul National University Bundang Hospital) ;
- Choi, Byeong-Don (Dept. of Radiation Oncology, Seoul National University Bundang Hospital) ;
- Park, Jin-Hong (Dept. of Radiation Oncology, Seoul National University Bundang Hospital)
- 방동완 (분당서울대학교병원 방사선종양학과) ;
- 석진용 (분당서울대학교병원 방사선종양학과) ;
- 정윤주 (분당서울대학교병원 방사선종양학과) ;
- 최병돈 (분당서울대학교병원 방사선종양학과) ;
- 박진홍 (분당서울대학교병원 방사선종양학과)
- Published : 2004.09.01
Abstract
Purpose : Film has been the primary tool in coincidence testing between the light field and the radiation field, which constitutes the quality assurance list of a linear accelerator. But there is a great chance of errors being different among the observer when using film. Thus this study set out to use the BIS(Beam Image System) in addition to film in comparing and evaluating coincidence results between the two fields and in searching for the improvement measures. Materials & Methods : Photon beam of 6 and 15MV was exposed to film and the BIS using a linear accelerator. The light and radiation fields were each
목적 : 선형가속기의 정도관리 항목 중 광조사야와 방사선조사야의 일치성 검사는 기존에 필름을 이용하여 측정하였다. 하지만 필름을 이용한 측정은 측정관찰자에 따라 오차가 크게 발생하여 이를 개선하기 위해 BIS(Beam Image System) 영상장치를 이용하여 필름과 조사야 일치성을 비교평가 하고자 하였다. 대상 및 방법 : 선형가속기를 이용하여 필름과 BIS 영상장치에 6, 15MV의 광자선을 조사하여 측정하였다. 조사야는 각각
Keywords