Verification of set-up accuracy using MLC leakage

MLC leakage를 이용할 Verification의 효용성 고찰

  • Yun InHa (Department of Radiation Oncology, Asan Medical Center) ;
  • Hong DongGi (Department of Radiation Oncology, Asan Medical Center) ;
  • Jung WangKwon (Department of Radiation Oncology, Asan Medical Center) ;
  • Seo DongRin (Department of Radiation Oncology, Asan Medical Center) ;
  • Yun HwaRyeong (Department of Radiation Oncology, Asan Medical Center) ;
  • Kim JungMan (Department of Radiation Oncology, Asan Medical Center)
  • 윤인하 (아산재단 서울아산병원 방사선종양학과) ;
  • 홍동기 (아산재단 서울아산병원 방사선종양학과) ;
  • 정왕권 (아산재단 서울아산병원 방사선종양학과) ;
  • 서동린 (아산재단 서울아산병원 방사선종양학과) ;
  • 윤화룡 (아산재단 서울아산병원 방사선종양학과) ;
  • 김정만 (아산재단 서울아산병원 방사선종양학과)
  • Published : 2004.03.01

Abstract

Purpose : Verification within 5mm is one of the important QA process of IMRP and SRS. Therefore, we improve accuracy of patients set-up using scale point. Materials and Methods : We compare MLC scale pointer with customerized port film graticule for patients who was underwent IMRT and SRS. Results : Scale pointer using MLC showed accurate location of patients landmark including divergency of beam, and any point of patient could be certified by means of cross of MLC. Conclusion : MLC scale pointer is effective and convenient method in verification of patients set-up using L-gram.

서론 및 목적 : 세기변조 방사선치료와 전신 정위적 방사선치료에서 5mm이내의 위치 정확성확인 (L-gram)은 환자치료 목적 달성을 위한 중요한 QA과정의 하나이다. 이에 본원에서 제작한 scale point를 이용하여 L-gram 영상을 얻어 환자치료의 정확성을 향상시키고자 한다. 대상 및 방법 : 본원의 MLC 부착장비인 CL2100C/D, CL21EX에서 치료중인 IMRT, SRS환자를 대상으로 기존의 상품화된 port film graticule과 본원에서 제작한 MLC scale pointer를 비교 분석하였다. 결과 : MLC를 이용한 scale pointer는 beam의 확산까지 고려한 정확한 위치를 나타내었고 MLC의 교차로 생기는 pixel 모양의 형성을 통해 어떠한 위치에서도 정확한 확인이 가능했다. 결론 및 고찰 : L-gram 확인시 약간의 시간적 투자($1{\sim}2$분)로서 정확한 위치를 나타냄으로서 L-gram 확인시간 단축과 치료부위의 정확한 위치 확인을 할 수 있었다.