DOI QR코드

DOI QR Code

Conceptual Design of Soft X-ray Microscopy for Live Biological Samples

  • Kim, Kyong-Woo (X-ray Microscopy Research Center, Wonkwang University) ;
  • Nam, Ki-Yong (X-ray Microscopy Research Center, Wonkwang University) ;
  • Kwon, Young-Man (X-ray Microscopy Research Center, Wonkwang University) ;
  • Shim, Seong-Taek (X-ray Microscopy Research Center, Wonkwang University) ;
  • Kim, Kyu-Gyeom (X-ray Microscopy Research Center, Wonkwang University) ;
  • Yoon, Kwon-Ha (Department of Radiology, X-ray Microscopy Research Center, Wonkwang University School of Medicine)
  • 투고 : 2003.09.25
  • 발행 : 2003.12.01

초록

This study describes the conceptual design of a soft x-ray microscope system based on a laserbased source for biomedical application with high resolution (${\leq}$50nm). The laboratory scale soft x-ray microscope consists of high power laser plasma x-ray source and grazing incidence mirrors with high reflectivity. The laser plasma source used for developing this system employs Q-switched Nd-YAG pulsed laser. The laser beam is focused on a tantalum (Ta) target. The Wolter type I mirror was used as condenser optics for sample illumination and as objective mirror for focusing on a detector. The fabrication of the Wolter type I mirror was direct internal cutting using ultraprecision DTM. A hydrated biological specimen was put between the two silicon wafers, the center of which was $Si_3N_4$ windows of 100㎚ thickness. The main issues in the future development work are to make a stable, reliable and reproducible x-ray microscope system.

키워드

참고문헌

  1. J. Kirz, C. Jacobsen, and M. Howells, 'Soft x-ray microscopes and their biological applications,' Quarterly Reviews of Biophysics, vol. 28, no. 1, pp. 33-130, 1995 https://doi.org/10.1017/S0033583500003139
  2. J. Thieme, J. Niemeyer, G. Machulla, and U. Schulte-Ebbert, 'Aggregation of colloids observed by X-ray microscopy,' X-Ray Microscopy and Spectromicroscopy, pp. II-11 - II-20, 1998
  3. V. H. Wolter, 'Spiegelsysteme streifenden Einfalls als abbildende Optiken fur Rontgenstrahlen,' Annalen der Physik, vol. 10, pp. 94 - 114, 1952 https://doi.org/10.1002/andp.19524450108
  4. G. Schmahl and D. Rudolph, 'Lichtstarke Zonenplatten als abbildende Systeme fur weiche Rontgenstrahlung (High power zone plates as image forming systems for soft x-rays),' Optik, vol. 29, pp. 577-585, 1969
  5. C. Jacobsen, 'Soft x-ray microscopy;' Trends in Cell Biology, vol. 9, pp. 44-47, 1999 https://doi.org/10.1016/S0962-8924(98)01424-X
  6. B. Niemann, D. Rudolph, G. Schmahl, M. Diehl, J. Thieme, W. Meyer-Ilse, W. Neff, R. Holz, R. Lebert, F. Richter, and G. Herziger, 'An x-ray microscope with a plasma x-ray source,' Optik; vol. 84, pp. 35-36, 1990
  7. D. Attwood, Soft X-rays and Extreme Ultraviolet Radiation, (Cambridge University Press, Cambridge, UK, 1999)
  8. K. H. Yoon, K. W. Kim, and Y. D. Shin, 'Design of laser plasma-based X-ray microscope system using grazing incidence mirror for biology application,' in VII International Conference on X-ray Microscopy 02, pp. 79, 2002
  9. http://www-cxro.lbl.gov

피인용 문헌

  1. Laboratory-size three-dimensional x-ray microscope with Wolter type I mirror optics and an electron-impact water window x-ray source vol.85, pp.9, 2014, https://doi.org/10.1063/1.4894468