Comparison of Air Kermabased and Absorbed Dose to Waterbased Protocols in the Dosimetry of High Energy Electron Beams
고 에너지 전자선에 대한 공기커마와 물 흡수선량에 기반한 프로토콜간의 비교
Abstract
A few years ago, a proposal was made to change the dosimetry from the air kerma-based reference dosimetry to the absorbed dose-based reference dosimetry for all radiotherapy beams of ionizing radiation to improve the accuracy of dosimetry. Here, we present a dosimetry study in which the two most widespread absorbed dosebased protocols (IAEA TRS398 and AAPM TG51) were compared with an air kermabased protocol (IAEA TRS-277) by measuring the absorbed dose in the same reference depth. Measurements were performed in three clinical electron beam energies using a PTW 30002 cylindrical chamber, and Markus and Roos planeparallel chambers.
흡수선량 측정의 정확성을 향상시키기 위하여 기존의 공기커마 혹은 조사선량 표준에서 물 흡수선량 표준으로 선량측정의 패러다임이 변화하고 있다. 본 연구는 기존의 프로토콜에서 물 흡수선량 기반의 프로토콜로의 전환을 모색하는 입장에서 현재 임상에서 널리 사용하고 있는 IAEA TRS277과 물 흡수선량 표준에 토대를 두고 있는 IAEA TRS398 및 AAPM TG51 프로토콜에 대하여 동일한 기준깊이에서 물 흡수선량을 측정하여 상호 비교하였다. Varian 2100 C/D 선형가속기로부터의 세 종류의 전자선 에너지에 대해 PTW 30002 원통형 이온함과 Markus 및 Roos 평행평판형 이온함을 사용하여 측정하였고