Cellular Force Sensing for Force Feedback-Based Biological Cell Injection

힘 피드백 기반의 세포조작을 위한 세포막 침습력 측정

  • 김덕호 (한국과학기술연구원 마이크로시스템센터) ;
  • 윤석 (한국과학기술연구원 마이크로시스템센터) ;
  • 강현재 (한국과학기술연구원 마이크로시스템센터) ;
  • 김병규 (한국과학기술연구원)
  • Published : 2003.12.01

Abstract

In biological cell manipulation, manual thrust or penetration of an injection pipette into an embryo cell is currently performed by a skilled operator, relying on visual feedback information only. Accurately measuring cellular forces is a requirement for minimally invasive cell injections. Moreover, the cellular force sensing is essential in investigating the biophysical properties for cell injury and membrane modeling studies. This paper presents cellular force measurements for the force feedback-based biomanipulation. Cellular force measurement system using piezoelectric polymer sensor is implemented to measure the penetration force of a zebrafish egg cell. First, measurement system setup and calibration are described. Second, the force feedback-based biomanipulation is experimentally carried out. Experimental results show that it successfully supplies real-time cellular force feedback to the operator at tens of uN and thus plays a main role in improving the reliability of biological cell injection tasks.

Keywords

References

  1. Tan, K. K. and Ng, S. C., 2001, 'Computer Controlled piezo Micromanipulation System for Biomedical applications,' Engineering Science and Education Journal, pp.249-256 https://doi.org/10.1049/esej:20010606
  2. Yanagida, K., Katayose, H., Yazawa, H.Y., Kimura, K., and Sato, A., 1998, 'The Usefulness of a Piezo-Micromanipulator inintracytoplasmic Sperm Injection in Humans,' Human Reproduction, Vol.14, No.2, pp.448-453 https://doi.org/10.1093/humrep/14.2.448
  3. Nakayama, T., Fujiwara, H., Tastumi, K., Fujita, K., Higuchi, T., and Mori, T., 1998, 'A New Assisted Hatching Technique Using a Piezo-Micromanipulator,' Fertility and Sterility, Vol.69, No.4 https://doi.org/10.1016/S0015-0282(98)00017-X
  4. Menciassi, A., Scalari, G., Eisinberg, A., Anticoli, C., Francabandiera, P., Carrozza, M. C., and Dario, P., 2001, 'An Instrumented Probe for Mechanical Characterization of Ssoft Tissues,' Biomedical Microdevices, pp. 149-156 https://doi.org/10.1023/A:1011454427384
  5. Kimura, Y., Yanagimachi, R., 1995, 'Intracytoplasmic Sperm Injection in the Mouse,' Biology of Reproduction, vol. 52, no. 4, pp. 709-720 https://doi.org/10.1095/biolreprod52.4.709
  6. Sun, Y., Nelson, B.J., 2001, 'Microrobotic Cell Injection,' in Proceedings of IEEE International Conference on Robotics and Automation, Seoul, Korea, Vol. 1, pp. 620-625 https://doi.org/10.1109/ROBOT.2001.932619
  7. Fung, C. K. M., Elhajj, I., Li, W. J., and Xi, N., 2002, 'A 2-D PVDF Force Sensing System for Micromanipulation and Microassembly,' in Proceedings of IEEE International Conference on Robotics and Automation, Washington DC, USA, pp. 1489-1494 https://doi.org/10.1109/ROBOT.2002.1014754
  8. Dargahi, J., 2000, 'A Piezoelectric Tactile Sensor with Three Sensing Elements for Robotic, Endoscopic and Prosthetic Applications,' Sensors and Actuators, Vol.80, pp.23-30 https://doi.org/10.1016/S0924-4247(99)00295-2
  9. Nam, Y. S., and Sasaki, M., 2002, 'Strain Rate Self-Sensing for a Cantilevered Piezoelectric Beam,' KSME International Journal, Vol.16, No.3, pp.310-319
  10. Lewis, W., Rosa, B., Peter, L., Elliot, M., Jim, S., and Thomas, M. J., 2002, Principles of development, Oxford University Press, 2nd edition