Abstract
This study was performed to prepare the verification film for localizing beam-target position with the Photon Knife radiosurgery system (PKRS) using linear accelerator(Mitsubishi, Model ML-15MDX). We developed a laser calibration system using a reticle of transparent lucite to detect Inlet and outlet beams. We verified fixation of the second collimator with film mounted on a holder in the shape of an octagon block 5cm apart from the isocenter. The film was exposed to photon beams of linear accelerator at an interval of 45 degrees during the gantry movement. There were no shifts in the beam of the second collimator during gantry movement. We used a position marker which is designed a head-shaped small lead block and a 10 mm in diameter of steel bead in the plastic tube. The position marker helped to verify the beam directions with patient position in multi-arc and trans-multi-arc of PKRS The verification of beam alignments showed an average 0.8
선형가속기(Mitsubishi, ML15MDX)를 이용한 방사선수술시스템인 Photon Knife에서 Linac-gram을 통해 선속-표적 위치를 확인하여 신뢰성 있는 시술을 유지하도록 하였다. 선속-레이저광 교정 기구를 제작하여 레이저광의 입사점과 사출점을 조사하여 빔의 위치결정에 이용하였다. 선형가속기에 부착한 보조 콜리메이터의 고정을 확인하기 위해 Isocenter에서 5 cm 떨어진 위치에 팔각형 필름지지체를 두도록 제작하고 확인용 필름(Kodak X-omat V2)을 설치하였다. 필름에 선형가속기의 지지체를 45