Clinical Use of Shielding Block Correction factors

차폐블록보정인자의 임상적 응용

  • Published : 2003.06.01

Abstract

In this study, we measured shielding block correction factors for irregular fields and compared them with published data for the square blocked field. We devised a methods to measure the factors at an arbitrary depth in phantom. The measurements were performed for 12 shielding blocks used in radiation therapy. The measured correction factors for irregular blocked fields were consistent within $\pm$0.5% with those of the square blocked fields. Our results show that the shielding block correction factors for the typical square blocked fields can be used in clinical dose calculations for irregular blocked fields. However, for small fields, we suggest that verification be done by measurement.

부정형 차폐조사면에 대하여 차폐블록보정인자를 측정하고 보고된 정방형 차폐조사면에 대한 결과들과 비교하여 이 인자의 임상적 응용을 논의하였다. 차폐블록보정인자를 팬텀속 임의의 깊이에서 측정할 수 있는 방법을 고안하고 차폐블록 12가지에 대한 측정을 수행하였다. 부정형 차폐조사면에 대한 차폐블록 보정인자는 정방형 차폐조사면에 대한 측정결과들과 $\pm$0.5% 이하의 차이를 보였다. 임상에서 일반적인 부정형 차폐조사면에 대한 선량계산시 정방형 차폐조사면에 대한 차폐블록보정인자를 적용할 수 있으나, 매우 작은 차폐조사면의 경우에는 실측을 통한 확인이 요구된다.

Keywords