Measurement and Characterization of Three Dimensional Luminous Flux

광원주위의 3차원 광선속 측정 및 평가

  • 최종운 (호남대 공대 광전자공학과) ;
  • 유문종 (호남대 공대 광전자공학과)
  • Published : 2003.07.01

Abstract

This paper describes the construction, measurement, and characterization of an instrument for the distribution of luminous flux. This is done by measuring the flux around a light source while a detector and a source is rotating, and integrating it over an entire imaginary surface surrounding the source. We make a gonio radiometer to realize the scales of total luminous flux and geometrically integrate to get total spectral flux from standards of irradiance and illuminance. The uncertainties of a total flux in the gonioradiometry are 1.3%, and 0.4% below the standard lamps of NIST and NIM for each other.

Keywords

References

  1. A. Russell Schaefer, Kshitij Mohan, 'A New Goniordiometer for Total Flux Measurements,' pp. 349-353, Journal of IES, July 1974
  2. G. Sauter, 'Goniophotometry: new calibration method and instrument design,' Metrologia, Vol. 32, pp. 685-688, 1995/96 https://doi.org/10.1088/0026-1394/32/6/58
  3. W. J. Brown, 'Realization of the NML Scale of Total Luminous Flux,' Metrologia Vol 15, pp 167-171, 1979 https://doi.org/10.1088/0026-1394/15/4/001