Development of Videooculograph for Vestibular Function Test

전정 기능 평가를 위한 영상 안구 운동 측정 시스템의 개발

  • 김수찬 (연세대학교 의과대학 의학공학교실) ;
  • 남기창 (연세대학교 의과대학 의학공학교실) ;
  • 이원선 (연세대학교 의과대학 이비인후과학교실) ;
  • 김덕원 (연세대학교 의과대학 의학공학교실)
  • Published : 2003.05.01

Abstract

Videooculography (VOG) is one of the eye-movement measurement methods used for objective evaluation of vestibule -ocular reflex. A key feature of VOG is to estimate accurately the center of pupil and ocular torsion with being less influenced by the upper eyelid droop, eyelashes, corneal reflection, and eye blinks. Especially, it Is important to find the accurate center of the pupil in 3-D VOG because the inaccurate pupil center causes significant errors on measuring torsional eye movement. A new algorithm was proposed to find the center of pupil which is a little influenced by factors mentioned above. In this study, real time three-dimensional VOG which can measure horizontal, vortical, torsional eye movements, and the diameter of pupil was implemented using the proposed method.

영상 안진기(VOG, Videooculograph)는 전정 안반사의 객관적 평가 방법으로 이용되고 있는 안구 운동 측정 방법 중의 하나이다. 영상 안진기의 핵심 기술은 눈꺼풀이나 속눈셉, 각막 반사, 그리고 눈 깜박임과 같은 영향에 덜 민감하면서 동공의 중점과 회전 운동을 정확하게 찾는 것이다. 회전 운동의 측정에 있어 부정확한 동공의 중점은 부정확한 회전 운동 측정으로 이어지므로 3차원 영상 안진기에서는 중점의 정확한 추정이 매우 중요하다. 이를 위해 잡음에 영향을 적게 받는 동공 중점을 찾는 새로운 알고리즘을 제안하고 알고리즘의 성능에 영향을 주는 인자에 대해 분석하여 최적의 조건을 찾았다. 그리고 제안한 방법을 이용하여 안구의 수평, 수직, 회전 운동, 그리고 동공의 크기 변화를 측정할 수 있는 실시간 3차원 VOG 시스템을 구현하였다.

Keywords

References

  1. I. R. J. Leigh and D. S. Zee, The Neurology of Eye Movements. 2nd ed. F.A. Davis. pp. 1-9, 1991
  2. D. W. J. Chung, M. Eizenman, B. S. K. Cheung, R. C. Frecker, 'Estimation of ocular torsion with dynamic changes in pupil size', in Proc. of 16th Annual International Conference of the IEEE. Baltimore, MD, USA, 1994 https://doi.org/10.1109/IEMBS.1994.415215
  3. M. Brigell, M. Bach, C. Barber, K. Kawasaki, A. Kooijman, 'Guidelines for calibration of stimulus and recording parameters used in clinical electrophysiology of vision', Calibration-Standard Committee of the International Society for Clinical Electrophysiology of Vision(ISCEV). Doc Ophthalmol, vol. 95, no. 1, pp. 1-14, 1998 https://doi.org/10.1023/A:1001724411607
  4. C. Fernandez and J. R Lindsay, 'Vestibular tests: methods of recording eye movements', Arch Otolaryngol, vol. 82, no. 6, pp. 664-6, 1965 https://doi.org/10.1001/archotol.1965.00760010666022
  5. A. M. Eckert and M. Gizzi, 'Video-oculography as part of the ENG battery', Br J Audiol, vol. 32, no. 6, pp. 411-6. 1998 https://doi.org/10.3109/03005364000000092
  6. D. Robinson, 'A method of measuring eye movement using a scleral search coil in a magnetic field', IEEE Trans Biomed Eng, vol. 10, pp. 137-145, 1963
  7. M. Hatamian and D.J. Anderson, 'Design consideration for a real-time ocular counter roll instrument', IEEE Trans Biomed Eng, vol. 30, no. 5, pp. 278-288, 1983 https://doi.org/10.1109/TBME.1983.325117
  8. A. Iijima, H. Minamitani, and N. Ishikawa, 'Image analysis of quick phase eye movements in nystagmus with high-speed video system', Med Biol Eng Comput, vol. 39, no.1, pp. 2-7, 200l https://doi.org/10.1007/BF02345259
  9. S. T. Moore, I. S. Curthoys, and S.G. McCoy, 'VTM-an image-processing system for measuring ocular torsion', Comput Methods Programs Biomed, vol. 35, no. 3, pp. 219-30, 1991 https://doi.org/10.1016/0169-2607(91)90124-C
  10. T. Haslwanter and S.T. Moore, 'A theoretical analysis of three-dimensional eye position measurement using polar cross-correlation', IEEE Trans Biomed Eng, vol. 42, no. 11, pp. 1053-61, 1995 https://doi.org/10.1109/10.469371
  11. D. Zhu, S.T. Moore, and T. Raphan, 'Robust pupil center detection using a curvature algorithm', Corrput Methods Programs Biomed, vol. 59, no. 3, pp. 145-57, 1999 https://doi.org/10.1016/S0169-2607(98)00105-9
  12. J. P. Ivins, J. Porrill, and J.P. Frisby, 'Deformable model of the human iris for measuring ocular torsion from video images', IEE Proceedings-Vision Image and Signal Processing, vol. 145, no. 3, pp. 213-220, 1998 https://doi.org/10.1049/ip-vis:19981914
  13. National Instruments, IMAQ Vision for G Reference Manual. Austin, Texas: National Instruments, 1997
  14. 김수찬, 남기창, 김덕원, 정운교, 이원상, '수직 안구 운동 측정을 향상시킨 실시간 video eye tracking', 대한의용생체공학회 춘계학술대회, vol. 24, no. 1, pp. 240-1, 2001
  15. S. C. Kim, D. W. Kim, K. C. Nam W. S. Lee, M. R. Bae, C. Y. Ryu, 'Minimization of Influence by the Partial Eyelid Closure in Measuring 3-D Eye Movements', 2nd Joint Conference of the IEEE Engineering in Medicine and Biology Society and the Biomedical Engineering Society, Houston, Texas USA, October, pp. 23-26, 2002
  16. K. J. Sung and D. J. Anderson. 'A video eye tracking system based on a statistical algorithm', Proceedings of the 36th Midwest Symposium of Circuits and Systems at Detroit, MI, USA 1993 https://doi.org/10.1109/MWSCAS.1993.342994