Development of Movable Detection System for Efficiency Measurement in 3-PM Liquid Scintillation Counting

  • Hwang, H-Y (Mokwon University) ;
  • Kwak, S.I (Mokwon University) ;
  • Cho, Y.H (Mokwon University) ;
  • Byun, J.I (Mokwon University) ;
  • Lee, H.Y (Mokwon University) ;
  • Seo, J.S (Mokwon University) ;
  • Kwak, J.Y (Mokwon University) ;
  • Lee, J.M (Korea Research Institute of Standards and Science) ;
  • Lee, K.B (Korea Research Institute of Standards and Science) ;
  • Park, T.S ;
  • Chung, K.H ;
  • Lee, C.W
  • 발행 : 2003.04.01

초록

We developed an improved 3-PM liquid scintillation counting (3-PM LSC) method in which three detectors can be displaced to back and forth directions, and a data acquisition system being able to provide the values for all parameters required for the method. The detectors are entirely located in a 20-mm lead chamber of an inner diameter of 500 mm. A saw-toothed gear ties up all detectors so as to move them uniformly, up to 50 mm with unit of 1 mm. The data acquisition system was designed in an integrated circuit to perform the necessary works such as fast amplification, discrimination, coincidence and logic analysis. It generates values of nine parameters among twelve's generated in the 3-PH LSC method. The dead time of each counting channel is of extending type, valving from 10 to 100 $mutextrm{s}$. We measured the TDCR values with an unquenched liquid scintillation source 1"C by displacing the detectors with a step of 2.5 mm away from counting vial. Their values were derived on the range from 0.9 to 0.6. The extent is three times wider than those regions observed by applying the defocalization technique.ique.

키워드

참고문헌

  1. Pochwalski, K, Broda, R. and Radoszewski, T. Sandardization of pure beta emitters by liquid scintillation counting. Appl. Radiat. Isot. 39, 165 (1988) https://doi.org/10.1016/0883-2889(88)90162-1
  2. Simpson B.R.S. and Meyer B.R. Activity measurement of 204T1 by Direct liquid scintillation methods. Nucl. Instr. and Meth. A 369, 340 (1996) https://doi.org/10.1016/S0168-9002(96)80005-2
  3. Hwang, H. Y., Park, T. S., Lee, J. M. and Han, K. H. Development of a three dimensional data acquisition method for sandardization of beta-emitting nuclides. Appl. Radiat. Isot. 52, 393 (2000) https://doi.org/10.1016/S0969-8043(99)00184-0
  4. R. Broda, K. Maleka, T. Terlikowska and P. Cassette Study of the influence of the Ls-cocktail composition for the standardization of radionuclides using the TDCR model. Appl. Radiat. Isot. 56, 285. (2002) https://doi.org/10.1016/S0969-8043(01)00202-0
  5. B.R.S. Simpson and B. R. Meyer Further investigations of the TDCR efficiency calculation technique for the direct determination of activity. Nucl. Instr. and Meth. A 312, 90 (1992) https://doi.org/10.1016/0168-9002(92)90134-P
  6. A.A. Noujaim, C. Ediss and L.I. Weibe Liquid Scintillation Science and technology Academic Press Inc. (1976)
  7. Michael J. Kessier Liquid Scintillation Analysis Science and technology Packard Instrument Co., Inc. (1989)
  8. International Commission on Radiation Units and Measurements, Particle Counting in Radioactivity Measurements, ICRU Report 52, pp. 27-36. (1994)
  9. R. Staubert, E. Bohm, K. Hein, K. Sauerland and J. Trmper Possible effect of photomultiplier afterpulses on scintillation counter measurements Nucl. Instr. and Meth. 84, 297 (1970) https://doi.org/10.1016/0029-554X(70)90276-4
  10. Broda, R., Pochwalski, K. and Radoszewski, T. Calculation of liquid scintillation detector efficiency. Appl. Radiat. Isot. 39, 159. (1988) https://doi.org/10.1016/0883-2889(88)90161-X
  11. A. G. Malonda and B.M. Coursey Calculation of beta particle counting efficiency for liquid scintillation systems with three phototubes App. Radiat. Isot. 39 1191 (1988) https://doi.org/10.1016/0883-2889(88)90098-6