Consideration of Surface Dose and Depth of Maximum Dose Using Various Detectors for High Energy X-rays

측정기에 따른 고에너지 X-선의 표면 선량 및 최대 선량 지점 고찰

  • Lee Yong Ha (Departments of Radiation Oncology Yonsei University Wonju College of Medicine) ;
  • Park Kyung Ran (Departments of Radiation Oncology Yonsei University Wonju College of Medicine) ;
  • Lee Jong Young (Departments of Radiation Oncology Yonsei University Wonju College of Medicine) ;
  • Lee Ik Jae (Departments of Radiation Oncology Yonsei University Wonju College of Medicine) ;
  • Park Young Woo (Departments of Science, Yonsei University Wonju College of Medicine) ;
  • Lee Kang Kyoo (Department of Radiation Oncology, Wonkwang University College of Medicine)
  • 이용하 (연세대학교 원주의과대학 방사선종양학교실) ;
  • 박경란 (연세대학교 원주의과대학 방사선종양학교실) ;
  • 이종영 (연세대학교 원주의과대학 방사선종양학교실) ;
  • 이익재 (연세대학교 원주의과대학 방사선종양학교실) ;
  • 박영우 (연세대학교 원주의과대학 기초과학교실) ;
  • 이강규 (원광대학교 의과대학 방사선종양학교실)
  • Published : 2003.12.01

Abstract

Purpose: It is difficult to exactly determine the surface dose and the dose distribution In buildup region of high energy X-rays by using the conventional ion chamber. The aim of this study Is to evaluate the accuracy of widely used dosimetry systems to measure the surface dose and the depth of maximum dose (d$_{max}$). Materials and Methods: We measured the percent depth dose (PDD) from the surface to the d$_{max}$ in either a water phantom or in a solid water phantom using TLD-100 chips, thimble type ion chamber, diode detector, diamond detector and Markus parallel plate ion chamber for 6 MV and 15 MV X-rays, 10$\times$10 cm$^{2}$, at SSD=100cm. We analysed the surface dose and the d$_{max}$. In order to verify the accuracy of the TLD data, we executed the Monte Carlo simulation for 5 MV X-ray beams. Results: The surface doses In 6 MV and IS MV X-rays were 29.31% and 23.36% ior Markus parallel plate ion chamber, 37.17$\%$ and 24.01$\%$ for TLD, 34.87$\%$ and 24.06$\%$ for diamond detector, 38.13$\%$ and 27.8$\%$ for diode detector, and 47.92$\%$ and 35.01$\%$ for thimble type ion chamber, respectively. in Monte Carlo simulation for 6 MV X-rays, the surface dose was 36.22$\%$, which Is similar to the 37.17$\%$ of the TLD measurement data. The d$_{max}$ In 6 WV and 15 MV X-rays was 14$\~$16 mm and 27$\~$29 mm, respectively. There was no significant difference in the d$_{max}$ among the detectors. Conclusion: There was a remarkable difference in the surface dose among the detectors. The Markus parallel plate chamber showed the most accurate result. The surface dose of the thimble ion chamber was 10$\%$ higher than that of other detectors. We suggest that the correction should be made when the surface dose of the thimble ion chamber Is used for the treatment planning ion the supeficial tumors. All the detectors used In our study showed no difference in the d$_{max}$.

목적 .: 고에너지 X-선의 표면 선량과 선량보강(build-up) 영역에서의 선량 분포는 일반적으로 방사선 계측에 사용되는 전리함 측정기로는 정확한 선량 분포를 얻기가 매우 어렵다. 본 연구는 고에너지 X-선 선량 계측에 보편적으로 사용되고 있는 여러 측정기를 이용하여 팬톰 표면에서의 흡수선량과 최대 선량 지점(d$_{max}$)을 측정하여 측정기 사이의 정확성을 비교 분석하고, 각 치료 기관에서 보편적으로 사용되는 측정기 중 표면 선량 측정에 적절한 측정장치를 제안하고 그 유용성을 제시하고자 한다. 대상 및 방법 : 본 실험에서는 6 MV와 IS MV X-선에 대해 조사면이 10$\times$10 cm$^{2}$, SSD=100 cm에서 TLD, 팀블형전리함(thimble type ion chamber), 다이오드 검출기, 다이아몬드 검출기와 Markus 평행판 전리함 등을 이용하여 심부선량백분율(percent depth dose: PDD)을 측정하여, 표면 선량(suface dose)과 최대 선량 지점(dnu)을 비교 분석하고, 또한 TLD 측정 시와 동일 조건으로 Monte Cario 계산을 실행하여 TLD의 측정 결과와 비교하였다. 결과: 6 WV와 IS MV X-선에 대해 Markus 평행판 전리함을 이용하여 측정한 표면 선량은 각각 29.31$\%$와 23.36$\%$으로 측정되었으며, TLD는 37.17$\%$와 24.06$\%$, 다이아몬드 검출기는 34.78$\%$와 24.06$\%$, 다이오드 검출기는 38.18$\%$와 27.8$\%$, 팀블형 전리함은 47.92$\%$와 36.06$\%$ 였으며, Monte Cario 계산에 의한 표면 선량 값은 S MV X-선에 대해 TLD 측정 시와 동일한 조건으로 팬톰 내에 가상적인 TLD를 삽입한 경우 36.22$\%$로 실제 측정값 37.17$\%$와 유사하였다. 최대 선량 지점의 깊이는 모든 측정기에서 6 MV X-선에 대하여 14$\~$16 mm, IS MV X-선에서는 27$\~$29 mm사이의 측정기에 따라 작은 차이를 보였다. 결론 : 표면 선량의 경우에는 측정기에 따라 현저한 차이를 보였으며 Markus 평행판 전리함이 사용된 측정기 중가장 정확한 결과를 보였고, 팀블형 전리함의 경우 다른 측정기에 비해 약 10$\%$ 이상 높은 선량을 보여 피부 표면에 가까이 위치한 종양에 대한 방사선 치료 계뵉 시에는 임상에서 가장 보편적으로 사용되고 있는 팀블형 전리함의 선량 값을 그대로 사용하기에는 많은 오류가 발생하므로 가능한 표면 선량 측정에 적절한 측정기를 선택하여 사용하거나 측정기 특성을 고려한 보정이 필요할 것으로 생각된다. 최대 선량 지점(d$_{max}$)의 결과는 모든 측정기에서 비슷한 결과를 나타내고 있어 본 실험에서 사용한 모든 측정기는 그 특성에 상관없이 최대 선량 지점 측정에 사용이 가능함을 알 수 있었다.

Keywords

References

  1. Khan FM. Treatment Planning III, In:William M, Passano III, eds. The physics of radiation therpy, 2nd ed., Maryland, Williams & Wilkims, 1994:323-332
  2. DuboisD,BiceW,BradfordB,etal. Moldabletissueequivalent bolus for high-energy photon and electron therapy. Med Phys 1996;23(9):1547-1549 https://doi.org/10.1118/1.597820
  3. Manson DJ, Velkey DE, Purdy JA, Oliver GD. Measurement of surface dose using build-up curve obtained with extrapolation chamber. Radiology 1975;115:473-474 https://doi.org/10.1148/115.2.473
  4. VelkleyDE,MansonDJ,PurdyJA,OliverGD. Buildup region of megavoltage photon radiation source. Med Phys 1975;2:14-19 https://doi.org/10.1118/1.594158
  5. NilsonB,Montelius A. Fluenceperturbation in photon beams under non-equilibrium condition. Med Phys 1986;13: 191-195 https://doi.org/10.1118/1.595895
  6. Gerbi BJ, Khan FM. Measurement of dose in the build up region using fixed-separation plane-parallel ionization chambers. Med Phys 1990;17:17-26 https://doi.org/10.1118/1.596522
  7. Pedersen K, Andersen TD, Olsen DR, et al. Sensitivity and stability of LiF thermoluminescence dosimeters. Med Dosim 1995;20(4):263-267 https://doi.org/10.1016/0958-3947(95)02004-7
  8. Kron T , B u t s o n M , H u n t F , e t a l. TLD extrapolation for skin dose determination in vivo. Radiotherpy and Oncology 1996;41:119-123 https://doi.org/10.1016/S0167-8140(96)01795-1
  9. Ferguson S, Ostwald P, Kron T, Denham J. Verification of surfacedoseonPatientundergoinglowtomediumenergy X-ray therapy. Medical dosimetry 1995;20(3):161-165 https://doi.org/10.1016/0958-3947(95)00009-L
  10. Tello VM, Tailor RC, Hanson WF. How water equivalent are water-equivalent solid materials for output calibration of photon and electron beams? Med Phys 1995;22(7):1177-1189 https://doi.org/10.1118/1.597613
  11. Kron T, MetcalfeP,WongT. Thermoluminescencedosimetry of therapeutic x-rays with Lif ribbonsandrods.PhysMedBiol 1993;38:833-845
  12. Kron T, ElliotA,WongT,ShowellG,ClubbB,MetcalfeP.X-ray surface dose measurement using TLD extrapolation. Med Phys 1993;20(3):703-711 https://doi.org/10.1118/1.597019
  13. Yu C, Luxton G. TLD dose measurement: A simplified accuratetechniqueforthedoserangfrom0.5cGyto1000 cGy. Med Phys 1999;26(6):1010-1016
  14. Khan FM, Moore VC, Levitt SH. Effect of various atomic number absorbers on skin dose for 10-MeV X-rays. Radiology 1973;109:209
  15. VahcYW,ParkKR,LeeYH,etal .Improvementof X-ray beam quality for treating cancerusingdoublefocuselectric field strings. Rad Oncol Invest 1999;7:382-389 https://doi.org/10.1002/(SICI)1520-6823(1999)7:6<382::AID-ROI9>3.0.CO;2-X
  16. DingGX.Energyspectra,angularspread,fluenceprofile and dose distribution of6and18MVphotonbeams:resultsof Monte Carlo simulations for Varian 2100EX accelerator. Phys Med Biol 2002;47:1025-1046 https://doi.org/10.1088/0031-9155/47/7/303
  17. MalatarasG,KappasC,LovelockDMJ .AMonteCarloapproach to electron contamination sources in the Saturne-25 and -41. Phys Med Biol 2001;46:2435-2446 https://doi.org/10.1088/0031-9155/46/9/312
  18. Vahc YW, Park KR, Kwon OH, et al. The measurements of energy and distribution of scattered electronintherapeutic X-ray beam. Kor J Med Phys 2002;13:25-32
  19. Yang KM, Suh HS. Surface and percent depth doses for multileaf collimator conjunction with conventional block. Kor J Med Phys 2002;13:62-68
  20. ICRP.1990recommendationoftheinternationalcommission on radiological protection. ICRP publication 60, Oxford, 1991: 149-153
  21. ICRU. Determination of dose equivalents resulting from external radiation source. ICRUReport 39, Bethesda,1985
  22. Butson MJ, Rozenfeld A, Mathur JN, Carolan M, Wong TPY, MetcalfePE. A newradiotherapysurfacedosedetector:The MOSFET. Med Phys 1996;23:655-658 https://doi.org/10.1118/1.597702
  23. Lin JP, Chu TC, Lin SY, Liu MT. Skin dose measurement by using ultra-thin TLDs. Applide Radiation and Isotopes 2001;55:383-391 https://doi.org/10.1016/S0969-8043(01)00082-3
  24. AridEGA,BurnsJE,DayMJ,etal.Centralaxisdepthdosedata for use in radiotherapy. British J Radioloy 1996; supplement 25:90-98
  25. Sixel KE, Podgorsak EB. Buildup region and depth of dose maximum of megavoltage x-ray beams. Med Phys 1994; 21(3):411-416 https://doi.org/10.1118/1.597305