Reproducibility of Applicator Position with High dose rate brachytherapy in uterine cervical cancer

자궁경부암 환자의 근접치료시 재현성 평가

  • Kim Jong-Hwa (Department of Radiation Oncology, Ajou University Hospital) ;
  • Son Jung-Hae (Department of Radiation Oncology, Ajou University Hospital) ;
  • Jung Chil (Department of Radiation Oncology, Ajou University Hospital) ;
  • Kim Mi-Hwa (Department of Radiation Oncology, Ajou University Hospital)
  • 김종화 (아주대의료원 치료방사선과) ;
  • 손정혜 (아주대의료원 치료방사선과) ;
  • 정칠 (아주대의료원 치료방사선과) ;
  • 김미화 (아주대의료원 치료방사선과)
  • Published : 2003.09.01

Abstract

I. Purpose Brachytherapy is the main component in treatment of patients with uterine cervical cancer. The reproducibility of applicator position in the same patient at repeated treatments was very important for accurate dose delivery. It was aimed to evaluate the change of applicator location between each high dose rate(HDR) brachytherapy insertion in the patients with uterine cervical cancer. II. Materials and Methods From January 1999 to October 2001, total 52 patients were treated with external beam radiotherapy and HDR brachytherapy (Microselectron, Nucletron). During six to seven times of brachytherapy, all patients had three treatment plans. From the orthogonal radiographs, we measured the following variables; height from upper border of pubic bone to os (HPO), distance from sacral promontory to tip of tandem (DST), distance from coccyx to os (DCO), distance from tip of right ovoid to os (DRO), distance from tip of left ovoid to os (DLO), and distance from center of the first tandem source to ovoid (DTO). To evaluate the reproducibility of applicator position, it was calculated the standard deviation of differences between three insertions for the 7 parameters in each patient. III. Results The ranges of standard deviations of interfractional differences for the variables were as follows. 1)HPO : $0{\sim}0.79cm$ 2)DST : $0{\sim}0.9cm$ 3)DCO : $0.06{\sim}0.76cm$ 4)DRO : $0{\sim}0.53cm$ 5)DLO : $0{\sim}0.45cm$ 6) DTO $0{\sim}0.36cm$ IV. Conclusions There was some change in applicator position on repeated implants in our study. But variation of the interfractional differences was minimal; in all parameters, there were less than 1 cm. We are continued to try for reducing the geometric variation between each procedure.

I. 목적 외부방사선 치료와 같이 근접치료 시에도 재현성을 평가하기 위해 bone의 위치와 tandem, ovoid 간의 거리를 측정하여 치료 setting의 재현성을 평가함으로써 자궁경부암 환자의 근접치료시 치료의 재현성을 평가하고자 하였다. II. 대상 및 방법 1999년부터 2001년 10월까지 본원을 내원해서 근접치료를 시행한 자궁경부암 환자중 tandem, ovoid을 사용한 52명을 대상으로 근접치료시 촬영한 film(환자1명당 3회촬영)을 가지고 측정하였다. Os을 기준으로 환자의 center축을 잡고 tandem, ovoid의 위치와 pubic symphysis, sacrum, coccyx등과의 거리를 각각 측정하여 치료 setting의 재현성을 평가하였다. III. 결과 환자 52명에 대한 tandem, ovoid의 위치와 pubic symphysis, sacrum, coccyx등과의 거리에 대한 표준 편차 결과는 다음과 같은 분포로 나타났다. 1) 각 tandem angle 의 표준편차는 $0^{\circ}{\sim}2.89^{\circ}$이다. 2) pubic symphysis에서 Os까지의 높이는 $0cm{\sim}0.79cm$ 3) sacrum에서 tandem tip까지의 거리는 $0cm{\sim}0.9cm$ 4) Os level에서 coccyx까지의 거리는 $0.06cm{\sim}0.76cm$ 5) ovoid tip에서 Os까지의 거리는 오른쪽은 $0cm{\sim}0.53cm$ 6) ovoid tip에서 Os까지의 거리는 왼쪽은 $0cm{\sim}0.45cm$ 7) 첫번째 source위치에서 ovoid사이의 거리는 $0cm{\sim}0.36cm$ 8) tandem에서 ovoid tip까지의 거리는 $0cm{\sim}0.31cm$ IV. 결론 자궁경부암 치료에 있어 근접치료는 중요한 치료 방법이다. 환자 치료간에 재현성을 평가해본 결과 대체로 만족 할 만한 수치를 보였으며 고선량율 근접치료에서 재현성 평가는 중요할 것이고 계속적으로 평가할 필요가 있을 것이다.

Keywords