Efficacy Evaluation of Magnetic P.

자기장을 활용한 통증치료기의 효력평가

  • Published : 2003.12.30

Abstract

The goal of this study is construction and efficacy evaluation of MPC-25 (Magnetic Pain Control-25), a medical device using magnetic stimulation. MPC-25 consists of a main body containing power supply and control module and a bed containing magnetic core and coil. In distinction from electric pain control medical devices, magnetic pain control system is non-contact, so the patients need not take off their clothes to be treated. High amplitude current pulses are applied to a magnetic coil and induce time varying magnetic field generating eddy current in a conductor like human body. Clinical efficacy test was performed in patients suffering from musculoskeletal pain of lumbar, shoulder and joint. Degree of pain before and after treatment for 30 minutes was compared using Visual Analogous Scale. As a result, 14 cases out of 20 showed decreased pain perception, so the rate of efficacy is 70%. Reduction of pain perception was statistically significant (P=0.001 by Wilcoxon Signed Rank Test).

Keywords