Bone height measurements of implant sites : Comparison of panoramic radiography and spiral computed tomography

임플란트 매식부의 고경 평가 : 파노라마 촬영법과 나선형 전산화 단층촬영법의 계측 비교

  • Cho Bong-Hae (Department of Oral and Maxillofacial Radiology, School of Dentistry, Pusan National University)
  • 조봉혜 (부산대학교 치과대학 구강악안면방사선학교실)
  • Published : 2002.06.01

Abstract

Purpose: To compare the bone height of implant sites measured using panoramic radiography and spiral CT. Materials and Methods : The available bone height was determined for 263 maxillary and mandibular implant sites in 59 patients. Distortion was calculated using the metal bar for the panoramic radiographs. Results: Significant differences in mean bone height between the two imaging modalities were found in maxillary and mandibular anterior regions (p<0.05). The mean difference in bone height recorded by the two techniques was smallest in the maxillary and mandibular molar areas (0.8 mm), and greatest in the mandibular anterior region (1.3 mm). With the exception of the mandibular anterior region, ninety percent of all the sites showed measurement differences within 2 mm. Conclusion: A safety margin of 2 to 3 mm is called for when utilizing panoramic radiography, otherwise additional imaging modality such as computed tomography is necessary to obtain accurate measurements.

Keywords