Optimization of PET Scan Time Using Phantom Studies

팬텀 영상을 이용한 PET 스캔시간의 최적화 연구

  • 정하규 (연세대학교 의과대학 BK21 의과학사업단, 방사선의과학연구소) ;
  • 김동현 (연세대학교 의과대학 BK21 의과학사업단, 방사선의과학연구소) ;
  • 정해조 (연세대학교 의과대학 방사선의과학연구소, 진단방사선과학교실) ;
  • 손혜경 (연세대학교 의과대학 BK21 의과학사업단, 방사선의과학연구소) ;
  • 홍순일 (연세대학교 의과대학 BK21 의과학사업단, 방사선의과학연구소) ;
  • 윤미진 (연세대학교 의과대학 방사선의과학연구소, 진단방사선과학교실) ;
  • 이종두 (연세대학교 의과대학 방사선의과학연구소, 진단방사선과학교실) ;
  • 김희중 (연세대학교 의과대학 BK21 의과학사업단, 방사선의과학연구소, 진단방사선과학교실)
  • Published : 2002.09.01

Abstract

The measured attenuation correction with transmission (Tx) scans produced quantitatively accurate images. However, it was not clear for optimal emission (Ex) and Tx scan time in PET imaging. This study was to evaluate acceptable Ex and Tx scan time by simulating clinical situations using various phantoms. Cylindrical and NEMA phantom were used for $^{18}$ F-PET scan using 2D protocol in GE Advance PETTM scanner. Cylindrical phantom was filled with 136 MBq 18F, and five regions of interests (ROI) were drawn on 23 slices. NEMA phantom had three inserts containing water, air and polytetrafluoro-ethylene (PTFE). Outside of these inserts were filled with 309 MBq of $^{18}$ F, and total 12 ROIs were drawn on 23 slices. Scans were carried out according to five Ex scan times: 2, 5, 10, 15, and 30 min, and nine Tx scan times: 2, 3, 4, 5, 7, 10, 15, 20, and 30 min. Images were reconstructed using measured attenuation correction, and ROI analyses were performed for all images, and mean, standard deviation (SD), coefficient of variation and percent errors were calculated. For cylindrical phantom study, ROI mean and SD were decreased as Ex and Tx time increased. Coefficients of variation were kept constant, when Tx was greater than 10 min. The amount of error decreased for the increment of Ex time from 10 min to 15 min was almost the same to that from 15 min to 30 min. In NEMA phantom Tx 15 min showed the lowest er개r level when the percent errors for three inserts were summed for all of the Ex times. This study suggested that Ex 15 min and Tx 15 min were acceptable as optimal scan time for the scanning protocol and the dose of radiopharmaceuticals used in these phantom study.

Keywords