Progress in Medical Physics (한국의학물리학회지:의학물리)
- Volume 13 Issue 2
- /
- Pages.62-68
- /
- 2002
- /
- 2508-4445(pISSN)
- /
- 2508-4453(eISSN)
Surface and Percentage Depth Doses for Multileaf Collimator Conjunction with Conventional Block
다엽 콜리메이터와 제작차폐물의 동시 사용시 표면선량 변화
Abstract
A muiltileaf collimator (MLC) is used as a replacement for conventional blocks. The MLC, however may not be appropriate for a fine field shaping. For the fine field shaping, conventional block can be added under the MLC. But it may significantly affect on the dosimetric characteristics such as surface dose of skin, buildup region and percent depth doses. We performed the study to evaluate the surface dose and the maximum depth dose using MLC conjunction with conventional blocks for various field sizes and energies. We confirmed the surface dose was increased by using the additional conventional block under the MLC ranging from 10 to 35.6% according to various field sizes and radiation beam energies. The surface dose was effectively reduced by application of 2 or 3 m thickness of lead plate as electron filter.
최근 설치된 선형가속기는 다엽 콜리메이터를 장착하고 있다. 기본적으로 다엽 콜리메이트의 사용은 기존의 차폐블럭을 대체하는 역할을 한다. 그러나 다엽 콜리메이터는 폭 1 cm의 콜리메이터가 각각 움직여서 방사선 조사 범위를 만들어 정교한 조사범위를 만들기 어려운 경우가 있을 수 있다. 따라서 조사야를 보다 정밀하게 만들기 위해 다엽 콜리메이터 아래 추가 차폐물을 사용하게 된다. 다엽 콜리메이터 아래 추가 차폐물을 사용할 경우 차폐물과 환자의 피부표면과의 거리가 짧아져 피부표면 선량이 증가하게 되며 최대 선량점(D