심장 내 이식형 축류 혈액펌프의 임펠러 최적화를 위한 용혈량 예측

Prediction of Hemolysis in Intra-Cardiac Axial Flow Blood Pumps for Optimization of the Impellers

  • 김동욱 (순천향대학교 정보기술공학부) ;
  • 발행 : 2002.09.01

초록

Low hemolysis is one of the key factors in the production of successful rotary blood pumps. It is, however, difficult to identify the areas where hemolysis occurs. Computational fluid dynamics(CFD) analysis enables the engineer to predict hemolysis on a computer Fluid dynamics in five different axial flow pumps was analyzed 3-dimensionally using CFD software. The impeller was rotated at a speed which supplied a flow of 5L/min at a pressure difference of 100mmHg. Changes in the turbulent kinetic energy along streamlines through the pumps were computed. Reynolds' shear stress( (equation omitted) ) was calculated using the turbulent kinetic energy. Hemolysis was evaluated based on Reynolds'shear stress and its exposure time(t) : dHb/Hb=3.62$\times$10$^{-5}$ $t^{0.785}$$\tau$$^{2.416}$ . Hemolysis of the pumps was measured in vitro using fresh bovine blood to which citrate phosphate dextrose was added to prevent clotting. A pump flow of 5L/min was maintained at a pressure difference of 100mmHg for 3h. The normalized index of hemolysis(NIH) as measured. Reynolds' shear stress was high behind the impellers. The measured NIH and the calculated hemolysis(dHb/Hb) shoed a good correlation; NIH=0.0003(dHb/Hb) (r=0.90, n=6) in the range of NIH between 0.003 and 1.1. CFD analysis can predict the in vitro results of hemolysis as well as the areas where hemolysis occurs.ysis occurs.

키워드

참고문헌

  1. http://www.heart.or.kr/ppt/98/00.pdf
  2. http://www.konos.go.kr/Left_Statistics.htm#1
  3. Nose Y, 'Long-term ex-vivo implantation (Longer than one year) of rotary blood pumps', Artificial Organs, Vol. 21, pp. 863, 1997 https://doi.org/10.1111/j.1525-1594.1997.tb00245.x
  4. Yada I, Golding L.R, Harasaki H, Jacobs G, Koike S, Yozu R, Saito N, Fujimoto LK, Snow J, Olsen E, Murabayashi S, Venkatesen VS, Kiraly R, Nose Y, 'Physiopathological studies of nonpulsatile blood flow in chronic models', Artificial Organs, Vol. 29, pp. 520-525, 1983
  5. D. W. Kim, Y.Mitamura, 'A Study of in-vitro Performances of the Intracardiac Axial Flow Pump', J. of KOSOMBE, Vol. 19, No.1, pp. 33-38, 1998
  6. Y Mitamura, H Nakamura, E. Okamoto, R. Yozu, S. Kawada, D.W Kim, 'Development of the Valve-Pump: Axial flow pump implanted at the heart valve position preserving the diseased heart muscle', Artificial Organs 23(6):566-571, 1999 https://doi.org/10.1046/j.1525-1594.1999.06404.x
  7. K. Naito, K. Mizuguchi, Y. Nose, 'The need for standardizing the index of hemolysis', Artificial Organs, Vo1.18, No.1, p.7, 1994 https://doi.org/10.1111/j.1525-1594.1994.tb03292.x
  8. L. J. Wurzunger and R. Opitz, 'Hematological Principles of Hemolysis and Thrombosis with Special Reference to Rotary Blood Pumps', Proc. Internal Workshop Rotary Blood Pumps, p.19, 1991
  9. L.J. Wurzinger, R. Opitz and H.Eckstein, 'Mechanical Bloodtrauma. An Overview, Angeiologie', Vol.38, No.3, pp.81-97, 1986
  10. M. Giersiepen, L.J. Wurzinger, R. Opitz, and H.Reul, 'Estimation of shear stress-related blood damage in heart valve prostheses - in vitro comparison of 25 aortic valves', Artificial Organs, Vol.13, pp.300-306, 1990
  11. M. Umezu, T. Fujimoto, Y Shiraishi, K. Iwasaki, K. Kondo, and M. Arita, 'Contribution of Mechanical Engineering to the Field of Artificial Organs', Inaugural Japan - Australia Cardiovascular Bioengineering Meeting (Sydney, Australia), pp.2-3, 1999
  12. R. Paul, F. Schugner, H. Reul, and G. Rau, 'Recent Findings on Flow Induced Blood Damage : Critical Shear Stresses and Exposure Times Obtained with a High Shear Couette-System', ISAO, p.680, 1999