Development of Adaptive Spatial Filter to Improve Noise Characteristics of PET Images

PET 영상의 잡음개선을 위한 적응적 공간 필터 개발

  • Woo, S. K. (Department of Nuclear Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Center for Clinical Research, Samsung Biomedical Research Institute, Department of Computer Engineering, Konkuk University) ;
  • Choi, Y. (Department of Nuclear Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Center for Clinical Research, Samsung Biomedical Research Institute) ;
  • Im, K. C. (Department of Nuclear Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Center for Clinical Research, Samsung Biomedical Research Institute) ;
  • Song, T. Y. (Department of Nuclear Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Center for Clinical Research, Samsung Biomedical Research Institute) ;
  • Jung, J. H. (Department of Nuclear Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Center for Clinical Research, Samsung Biomedical Research Institute) ;
  • Lee, K. H. (Department of Nuclear Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Center for Clinical Research, Samsung Biomedical Research Institute) ;
  • Kim, S. E. (Department of Nuclear Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Center for Clinical Research, Samsung Biomedical Research Institute) ;
  • Choe, Y. S. (Department of Nuclear Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Center for Clinical Research, Samsung Biomedical Research Institute) ;
  • Park, C. C. (Department of Computer Engineering, Konkuk University) ;
  • Kim, B. T. (Department of Nuclear Medicine, Samsung Medical Center, Sungkyunkwan University School of Medicine, Center for Clinical Research, Samsung Biomedical Research Institute)
  • 우상근 (성균관대학교 의과대학, 삼성서울병원 핵의학과, 삼성생명과학연구소 임상의학연구센터, 건국대학교 컴퓨터.정보통신공학과) ;
  • 최용 (성균관대학교 의과대학, 삼성서울병원 핵의학과, 삼성생명과학연구소 임상의학연구센터) ;
  • 임기천 (성균관대학교 의과대학, 삼성서울병원 핵의학과, 삼성생명과학연구소 임상의학연구센터) ;
  • 송태용 (성균관대학교 의과대학, 삼성서울병원 핵의학과, 삼성생명과학연구소 임상의학연구센터) ;
  • 정진호 (성균관대학교 의과대학, 삼성서울병원 핵의학과, 삼성생명과학연구소 임상의학연구센터) ;
  • 이경한 (성균관대학교 의과대학, 삼성서울병원 핵의학과, 삼성생명과학연구소 임상의학연구센터) ;
  • 김상은 (성균관대학교 의과대학, 삼성서울병원 핵의학과, 삼성생명과학연구소 임상의학연구센터) ;
  • 최연성 (성균관대학교 의과대학, 삼성서울병원 핵의학과, 삼성생명과학연구소 임상의학연구센터) ;
  • 박장춘 (건국대학교 컴퓨터.정보통신공학과) ;
  • 김병태 (성균관대학교 의과대학, 삼성서울병원 핵의학과, 삼성생명과학연구소 임상의학연구센터)
  • Published : 2002.06.01

Abstract

A spatially adaptive falter was formulated to imrove PET image qualify and the Performance of the filter was evaluated using simulation and phantom and human PET studies. In the proposed filter. if a pixel was identified as the edge Pixel, the Pixel value was Preserved. Otherwise a Pixel was replaced by the mean of the pixel values weighted by 2:7: 2. A Pixel was identified as the edge Pixel. if it satisfies the following conditions : the number of ADs (absolute difference between center and neighborhood pixels) which is smaller than THl (($pix_max{\times}0.1/log_2(NPM)$, NPM : mean of 6 neighborhood pixels excluding minimum and maximum) is 8-k and the number of ADs which is lager than TH2 ($NPM{\times}0.1$) is k. where k : 2, 3, …, 6. The results of this study demonstrate the superior performance of the Proposed titter compared to Gaussian fitter, weight median filter and subset averaged median filter. The proposed tittering method is simple but effective in increasing uniformity and contrast with minimal degradation of spatial resolution of PET images and thus. is expected to Provide improved diagnositc quality PET images .

양전자방출단층촬영기(Positron emission tomography, PET) 영상 개선을 위하여 적응적 공간 필터를 개발하였으며. 개발한 필터의 성능을 시뮬레이션데이터. 모형 PET 영상과 환자 PET 영상을 이용하여 평가하였다. 경계화소로 검출된 화소와 윈도우내의 모든 화소값이 동일한 화소는 보존하고. 그 외 화소에 대하여 2:7:2 비율로 가중치를 주어 정열한 후 중앙의 9개 화소에 대한 평균값으로 대치하였다. 경계화소를 검출하기 위하여 두 개의 임계값(TH7, TH2)을 이용하였으며. 다음의 조건을 만족하면 경계화소로 판단하였다 . THl ($pix_max{\times}0.1/log_2(NPM)$, NPM :주변값중 최상위 값과 최하위 값을 제외한 주변값들의 평균) 보다 작은 ADs (중앙값과 주변값의 차에 대한 절대값) 개수는 8-k이고. TH2 ($NPM{\times}0.1$) 보다 큰 ADs 개수는 k. 여기서 k는 2, 3 ‥‥ 6의 값을 가진다 성능평가 결과 이 연구에서 제안한 필터가 가우시안 필터, 가중메디안 필터, 부분집합평균메디안 필터 등과 비교하여 우수한 성능을 제공하는 것을 관찰하였다. 본 논문에서 개발한 간단한 적응적 공간 필터는 공간 분해능 저하는 최소화하면서 균일도와 대조도를 향상시키는데 효과적 이여서 정확한 PET 영상 해석에 기여할 것으로 기대된다

Keywords

References

  1. Physics & Techonogy v.8 no.7-8 PET: 생체기능을 측정하는 영상기기 J.H.Kim;K.C,Im;Y,Choi
  2. Phsics in Nuclear Medicine J.A,Sorenson;M.E,Phelps
  3. J.Biomed Eng Res. v.21 no.5 A study on the ordered subsets expectation maximization reconstruction method using Gibbs priors for emission computed tomography K.C.Im;Y.Choi;J.H.Kim;S.J.Leee;S.K.Woo;H,K.Seo;K.H.Lee;S.E.Kim;Y.S.Choe;C.C.Park;B,T,Kim
  4. Phys Med Biol. v.32 The noise spectrum of CT images M.F.Kijewski;P.F.Judy
  5. CVGIP v.54 no.2 Performance evaluation of some noise reduction methods W.Y.Wu;M.J.J.Wang;C.M.Liu
  6. Digital Image Processing R.C.Gonzalea.;R.E.Woods
  7. Digital Image Processing Algorithms I.Prtas
  8. Digital Image Processing B,Jahne
  9. IEEE Trans.Acoust.Speech.Signal Pricessing v.ASSP-30 no.5 Median filter: some modifications and their properties T,A,Nodes;N,C,Gallagher;J,R.
  10. IEEE Trans,Circuits and Systems v.43 no.3 Weighted median filters: a tutorial L.Yin;R.Yang;M.Gabbonj;Y,Neavo https://doi.org/10.1109/82.486465
  11. Elsevier Science Publishers 2-D subset averaged median filters R,Suoranta;K.P.Estola
  12. Ann of Biomed Eng. v.24(suppl.1):S 62 An anthropomorphic mathematical cardiac torso(MCAT) phantom for emission computed tomofraphy research K.J.LaCroix;B,M.W.Tsui
  13. Korean J Nucl Med. v.30 Performance measurements of positron emission tomography: an investigation using General Electric Advance J,R,Lee;Y.Choi;Y,s,Choe;K.H.Lee;S,E,Kim;S.A.Shin;B.T.Kim
  14. Machine Vision R.Jain;R.Kasturi;B.G.Schunck