Validation of the Assay for the Glucose of Human Venous Blood Using a Handy Blood Sugar Measuring Instrument
간이혈당측정기를 이용한 정맥 혈당 측정의 타당성 검증 - 간이혈당측정기의 타당성 검증 -
- Hong, Kyoung-Sup (Clinical Trial Center, Seoul National University Hospital, Department of Pharmacology, Seoul National University, College of Medicine) ;
-
Bae, Kyun-Seop
(Clinical Research Support Center, Asan Medical Center) ;
-
Lim, Hyeong-Seok
(Clinical Trial Center, Seoul National University Hospital, Department of Pharmacology, Seoul National University, College of Medicine) ;
-
Chung, Jae-Yong
(Clinical Trial Center, Seoul National University Hospital, Department of Pharmacology, Seoul National University, College of Medicine) ;
- Kim, Myo-Kyoung (Department of Pharmacology, Seoul National University, College of Medicine) ;
-
Baik, Jung-Mi
(Clinical Trial Center, Seoul National University Hospital) ;
- Kang, Kyoung-Ja (Clinical Trial Center, Seoul National University Hospital) ;
- Park, Gui-Ju (Clinical Trial Center, Seoul National University Hospital) ;
- Shin, Sang-Goo (Clinical Trial Center, Seoul National University Hospital, Department of Pharmacology, Seoul National University, College of Medicine) ;
-
Jang, In-Jin
(Clinical Trial Center, Seoul National University Hospital, Department of Pharmacology, Seoul National University, College of Medicine)
- 홍경섭 (서울대학교병원 임상시험센터, 서울대학교 의과대학 약리학교실) ;
-
배균섭
(서울아산병원 임상연구지원센터) ;
-
임형석
(서울대학교병원 임상시험센터, 서울대학교 의과대학 약리학교실) ;
-
정재용
(서울대학교병원 임상시험센터, 서울대학교 의과대학 약리학교실) ;
- 김묘경 (서울대학교 의과대학 약리학교실) ;
-
백정미
(서울대학교병원 임상시험센터) ;
- 강경자 (서울대학교병원 임상시험센터) ;
- 박귀주 (서울대학교병원 임상시험센터) ;
- 신상구 (서울대학교병원 임상시험센터, 서울대학교 의과대학 약리학교실) ;
-
장인진
(서울대학교병원 임상시험센터, 서울대학교 의과대학 약리학교실)
- Published : 2002.06.30
Abstract
Background : Measurement of blood glucose is a basic part of patient care. Increasingly, blood glucose measurement has become a part of self-care. Devices featuring convenience and quickness are used to measure blood glucose and are used in settings often unsupervised by knowledgeable health care personnel. In this context, a validation test of a handy assurances that the devices that they are using or purchasing are capable of producing valid results. Method : In this study, we fulfilled requirements from the ICH Q2 guidelines for assurance of accuracy, precision, linearity and range. From the two subjects, blood glucose was measured repeatedly at post-prandial 1h/2h/3h/4h, 12 times each, by two investigators and with two instruments. Then, we compared the results of a handy blood sugar measuring instrument with those of a chemistry analyzer, Hitachi 7070 auto-analyzer, in a core laboratory in Clinical Trial Center, Seoul National University Hospital. Results : Mean accuracy, intra-assay precision and intermediate precision were 92.6%, 2.3%(min. 1.07%, max. 4.51%), 1.73%(min. 0%, max. 5.65%) each. In scatter plot of the results of a handy blood sugar measuring instrument and Hitachi 7070 auto-analyzer, linearity was apparent and the correlation coefficient(r),