The thickness of Cu Filter to reduce 1/2 of the patient dose

X선 진단시 피폭선량을 반으로 줄이기 위한 Cu Filter의 두께

  • 김정민 (고려대학교 보건대학 방사선과) ;
  • 김성철 (가천길대학 방사선과)
  • Published : 2001.06.30

Abstract

Medical X-ray examination are increased double for the last $6{\sim}8$ years. Therefore a patient exposure dose should be decrease half every 7 years. We made an experiment on copper filter thickness to decrease a patient exposure dose up to half and compared to the Image quality by MTF. The results as follow 1. A thin region like extremities needs a thicker Cu filter as compared a thick region. 2. 1/2 reduction filter must be thicker when kVp Increase. 3. Exposure factor should be increas when using 1/2 reduction filter ; extremity is 4.0 times, chest 2.9 times, skull 1.62 times, and abdomen 1.58 times 4. The MTF of using 1/2 reduction filter is lower than without filter. But no difference of visual image. 5. 1/2 reduction filter compared with double speed screen showed almost same image quality.