Characteristics of dose distribution for virtual wedge

가변형 쐐기필터의 선량분포에 관한 특성

  • 김부길 (전북대학교 의과대학 의공학교실) ;
  • 김진기 (전북대학교병원 치료방사선과)
  • Published : 2001.09.01

Abstract

We was investigate the dosimetric characteristics of the virtual wedge and it compared to the conventional fixed wedge. Also we was evaluate the quality factor of the experimental multi-channel dosimetry system for virtual wedge. Recently virtual wedge technique and wedge fraction methods are available through the computer controlled asymmetric collimator or the independent jaw in medical linear accelerator for radiation therapy. The dosimetric characteristics are interpreted by radiation field analyzer RFA-7 system and PTW-UNIDOS system. Experimental multi-channel dosimetry system for virtual wedge was consists of the electrometer, the solid detector and array phantom. The solid detectors were constructed using commercially diodes for the assessment of quality assurance in radiotherapy. And it was used for the point dose measuring and field size scanning. The semiconductor detector and ion chamber were positioned at a dmax, 5 cm, 10 cm, 20 cm depth and its specific ratio was determined using a scanning data. Wedge angles in fixed and virtual type are compared with measurements in water phantom and it is shown that the wedge angle 15$^{\circ}$, 30$^{\circ}$, 45$^{\circ}$were agree within 1$^{\circ}$ degree in 6, 10 MV photon beams. In PDD and beam flatness, experimental multi-channel disimetry system was capable of reproduceing the measured values usually to within $\pm$2.1% the statistical uncertainties of the data. It was used to describe dosimetric characteristics of virtual wedge in clinical photon beams. Also we was evaluate optimal use of the virtual wedge and improve the quality factor of the experimental multi-channel dosimetry system for virtual wedge.

방사선을 이용한 치료기인 의료용 선형가속기에 부착하여 선량분포 조절에 사용하는 가변형 쐐기필터(virtual wedge)의 선량분포 특성를 분석하고 쐐기각의 자료를 비교하여 가변형 쐐기필터 적용에서 오는 선량분포의 변화를 알아보고자 하였다. 기준전리함인 단일채널 PTW-UNIDOS과 3차원 선량계측시스템인 RFA-7 그리고 실험적으로 구성된 다중채널 검출기의 특성실험을 통하여 선량분포도 및 쐐기각을 구하고 이를 이용하여 실험적 검출시스템의 측정인자를 보정하였다. 가변형 쐐기필터의 쐐기각은 15$^{\circ}$, 30$^{\circ}$, 45$^{\circ}$ 에서 고정형 쐐기필터에 비교하여 1$^{\circ}$ 이내로 안정하였고, 60$^{\circ}$ 경우는 6 MV에서 1.2$^{\circ}$, 10WV 에서 2.2$^{\circ}$ 의 변화를 보였다. 측정에너지 영역 6, 10 MV 의료용 광자선을 적용하여 선량백분율과 빔평탄도 측정결과 검출기의 신뢰도가 평균 $\pm$2.1 % 이내였다. 고정형 쐐기필터를 이용한 쐐기각 변화율방법 적용시 방사선치료 과정에서 나타나는 선량오차 및 정상조직의 피폭을 줄이고 장비의 과도한 부하를 감소시키며 가변형 쐐기필터의 다양한 치료방법의 적용을 모색할 수 있었다. 가변형 쐐기필터 이용과 측정된 선량분포의 특성 분석으로 기존 고정형 쐐기필터 이용방법 및 기능을 보완하였고 실험적으로 구성된 계측시스템 측정장비의 대체방법 보정으로 가변형 쐐기필터의 다양한 치료방법적용과 보조물 사용에 따른 정확성과 효율을 높이는데 유용하리라 사료된다.

Keywords