용량형진공계 교정 불확도

Calibration uncertainty of a capacitance diaphragm gauge

  • 발행 : 2001.07.01

초록

1,333 Pa 용량의 용량형진공계를 저진공 국가표준기인 초음파간섭 수은주압력계(ultrasonic interferometer manometer)를 이용하여 교정하였다. 이 결과의 불확도를 국제표준화기구의 권고안의 A형 불확도, B형 불확도, 합성표준불확도, 확장불확도 등으로 구분하여 평가하였다. 표준압력 1.76 Pa~1.31$\times$$10^3$Pa에서의 합성표준불확도는 $1.38 \times10^{-2}\; Pa\sim3.03 \times10^{-1} $Pa이었으며, 상대불확도(합성표준불확도/표준압력)는 $2.3 \times 10^{-4}\;Pa\sim7.9 X\times10^{-3} $Pa 이었다.

We calibrated a capacitance diaphragm gauge(CDG) of 1,333 Pa range by using ultrasonic interferometer manometer (UIM) that was a national low vacuum standards system. And its uncertainties were evaluated according to International Organization for Standardization(ISO), they were named to A type uncertainty, B type uncertainty, combined standard uncertainty, and expanded uncertainty, We obtained that the combined standard uncertainties were $1.38 \times10^{-2}\; Pa\sim3.03 \times10^{-1} $Pa and the relative uncertainties(combined standard uncertainty/standard pressure) were $2.3 \times 10^{-4}\;Pa\sim7.9 X\times10^{-3} $Pa for this 1,333 Pa CDG.

키워드

참고문헌

  1. Guide to the Expression of Uncertainty in Measurement ISO
  2. 측정불확도 표현 지침 한국표준과학연구원
  3. Protocol of CCM International Key Comparison in Absolute Pressure NIST
  4. KSRI-87-46-IR 간섭원리에 의한 액주형 압력계 개발 정광화(외)
  5. KRISS-96-021-IR 초음파간섭 수은주 압력계의 정밀정확도 향상 연구 정광화(외)
  6. 한국진공학회지 v.5 no.3 홍승수(외)
  7. Draft A report: CCM Key Comparison CCM.P-K4 in Absolute Pressure NIST
  8. KRISS-C-09-006 용량형진공계의 교정절차 홍승수(외)