32-Channel Bioimpedance Measurement System for the Detection of Anomalies with Different Resistivity Values

저항률이 다른 내부 물체의 검출을 위한 32-채널 생체 임피던스 측정 시스템

  • 조영구 (경희대학교 동서의학대학원) ;
  • 우응제 (경희대학교 전자정보학부 동서의료공학과)
  • Published : 2001.12.01

Abstract

In this paper. we describe a 32-channel bioimpedance measurement system It consists of 32 independent constant current sources of 50 kHz sinusoid. The amplitude of each current source can be adjusted using a 12-bit MDAC. After we applied a pattern of injection currents through 32 current injection electrodes. we measured induced boundary voltages using a variable-gain narrow-band instrumentation amplifier. a Phase-sensitive demodulator. and a 12-bit ADC. The system is interfaced to a PC for the control and data acquisition. We used the system to detect anomalies with different resistivity values in a saline Phantom with 290mm diameter The accuracy of the developed system was estimated as 2.42% and we found that anomalies larger than 8mm in diameter can be detected. We Plan to improve the accuracy by using a digital oscillator improved current sources by feedback control, Phase-sensitive A/D conversion. etc. to detect anomalies smaller than 1mm in diameter.

인체의 각 조직은 서로 다른 저항률(resistivity)을 가지고 있고. 심장의 박동이나 호흡과 같은 생리현상은 해당 생체조직의 임피던스를 변화시킨다. 본 논문에서는 인체 내부에 존재하는 비정상 조직의 크기와 위치를 검출하기 위한 32-채널 생체 임피던스 측정 시스템에 대하여 기술한다. 이러한 기술은 유방암 조직의 경우와 같이 배경 조직과는 저항률이 다른 비정상 조직을 검출하는 경우에 응용할 수 있을 것으로 기대한다. 32-채널 생체 임피던스 측정 시스템을 위하여 32개의 복합형 전극과 32 채널의 정전류원을 사용하였다. 임피던스의 측정을 위해 50kHz의 정현파 전류를 주입하고. 유기되는 전압을 가변 이득 협대역 계측용 증폭기로 측정하고, 그 크기를 위상감응복조기로 검출하였다. 검출된 임피던스 신호는 A/D 변환하여 PC에 입력하였다. 전해질 팬텀을 이용한 실험에서 전체 시스템의 정확도는 2.42%이며, 직경 270mm인 팬텀 내부에 존재하는 직경 8mm 이상인 물체의 크기와 위치를 검출할 수 있었다. 본 연구의 결과를 기초로 다채널 생체 임피던스 측정 시스템의 정확도를 개선하여. 직경 lmm 이내의 물체를 검출하는 것이 향후의 연구 목표이다 이러한 정확도를 가지는 생체 임피던스 계측 시스템을 개발하면. 인체 내부의 임피던스 분포를 측정하는 EIT(electrical impedance tomography) 시스템과, 최근에 연구되고 있는 자기공명 임피던스 단층촬영(MREIT, magnetic resonance electrical impedance tomography)에도 응용이 가능할 것이다.

Keywords

References

  1. Electrical Impedance Tomography J.G. Webster(ed.)
  2. J. Med. Eng. Tech. v.21 no.6 Imaging with electricity;report of the European concerted action on impedance tomography K. Boone;D. Barber;B. Brown
  3. IEEE Trans. Biomed. Eng. v.34 no.11 Comparing reconstruction algorithms for electrical impendance tomography T.J. Yorkey;J.G. Webster;W.J. Tompkins
  4. Int. J. Imaging Sys. and Tech. v.2 NOSER;An algorithm for solving the inverse conductivity problem M. Cheney;D. Isaacson;J.C. Newell;S. Simske;J. Goble
  5. IEEE Trans. Med. Img. v.12 no.2 A robust image reconstruction in electrical impedance tomography E.J. Woo;P. Hua;J.G. Webster;W.J. Tompkins
  6. IEEE Trans. Med. Img. v.16 no.5 Regularized reconstruction in electrical impedance tomography using a variance uniformization constraint C. Cohen-Bacrie;R. Guardo
  7. IEEE. Trans. Biomed. Eng. v.45 no.7 An iterative Newton-Raphson method to solve the inverse conductivity problem P. Edic;D. Isaacson;G. Saulnier;H. Jain;J.C. Newell
  8. IEEE Trans. Med. Img. v.17 no.2 Tikhonov regularization and prior information in electrical impedance tomography M. Vauhkonen;D. Vadasz;P.A. Karjalainen;E. Somersalo;J. P. Kaipio
  9. IEEE Trans. Biomed. Eng. v.48 no.2 An image reconstruction algorithm for three-dimensional electrical impedance tomography A.L. Hyaric;M.K. Pidcock
  10. Inverse Problem v.17 Total size estimation and identification of multiple anomalies in the inverse electrical impedance tomography O. Kwon;J.K. Seo
  11. Comm. Korean Math. Soc. v.16 no.3 Electrical impedance imaging for searching anomalies O. Kwon;J.K. Seo;E.J. Woo;J.R. Yoon
  12. Safe Current Limits for Electromedical Apparatus (ANSI/AAMI ESI) AAMI
  13. Bioimpedance and Bioelectricity Basics S. Grimnes;O.G. Martinsen
  14. IEEE Trans. Biomed. Eng. v.40 no.1 Using compound electrodes in electrical impedance tomography P. Hua;E.J. Woo;J.G. Webster;W.J. Tompkins
  15. Medical Instrumentation: Application and Design(3rd. ed.) J.G. Webster(ed.)
  16. Design with Operational Amplifiers and Analog Intergrated Circuits(2nd ed.) S. Franco
  17. IEEE Trans. Med. Img. v.5 Distinguishability of conductivities by electric current computed tomography D. Isaacson
  18. IEEE Trans. Biomed. Eng. Detection of anomalies with different conductivities using electrical impedance tomography O. Kwon;J.R. Yoon;J.K. Seo;E.J. Woo;Y.G. Cho
  19. IEEE Trans. Biomed. Eng. Magnetic resonance electrical impedance tomography (MREIT);simulation study of J-substitution algorithm O. Kwon;E.J. Woo;J.R. Yoon;J.K. Seo
  20. (submitted to) IEEE Trans. Med. Img. J-substitution algorithm in magnetic resonance electrical impedance tomography (MREIT);phantom experiments for static resistivity images H.S. Khang;B.I. Lee;S.H. Oh;E.J. Woo;S.Y. Lee;M.H. Cho;O. Kwon;J.R. Yoon;J.K. Seo