Interlaboratory Comparison of Critical Current Measurements on Ag-sheathed Bi-2223 tapes

Bi-2223선재의 임계전류 측정기술 비교

  • Lee, Kyu-Won (Korea Research Institute of Standards and Science, Taejon) ;
  • Han, Gi-Youl (Korea Research Institute of Standards and Science, Taejon, Korea, Chungbuk National University, Cheongju, Chungbuk)
  • Published : 2001.01.01

Abstract

We have conducted two runs of interlaboratory comparison on Ag-sheathed Bi-2223 tapes to evaluate the level of measurement techniques for the critical current measurement. Two classes of specimens were prepared for parallel and serial routings and sent to four participating laboratories. The critical currents of specimens were measured at 77 K in zero magnetic field. In the first comparison, we used twenty different Bi-2223 tapes as specimens for comparison and participating laboratories measured the specimens using their own instruments and procedures. As a result, the scattering of data on the first comparison showed -3.0% to +l2.2% for the parallel routing and -0.7% to +l5.1% for the serial routing. Major sources of these variations were attributed to different measurement techniques. Thus, the second comparison of measurement was done on the same specimens under specified measurement conditions, particularly in terms of cooling procedure and sweep rate of the test current. The variations for the second comparison were decreased -3.1% to +3.2% far the parallel routing and -1.8% to +7.7% fur the serial routing.