Accuracy and Precision according to Optimal Preparation by Analysis of Urine Lead

뇨중 연 분석의 정확성 및 정밀성에 미치는 최적 전처리법

  • 장봉기 (순천향대학교 환경보건학과) ;
  • 정은희 (순천향대학교 환경보건학과) ;
  • 박종안 (순천향대학교 환경보건학과) ;
  • 손부순 (순천향대학교 환경보건학과) ;
  • 이종화 (순천향대학교 환경보건학과)
  • Published : 2001.03.01

Abstract

In order to propose a optimum method increasing accuracy and reproducibility in analysis of urine lead levels, we compared matrix modifier addition method and microwave digestion method. The results were as follows ; 1. Where the concentration of $Pd(NO_3)_2$(matrix modifier) was $50mg/{\ell}$, and ashing temperature was $900^{\circ}C$, the results were optimum. In case of matrix modifier addition method, the average coefficient of variation (CV) of normal man was 24.1%, and lead worker was 7.9%. The average urine lead levels of standard were $10.42\mu{g}/{\ell}$ and $19.89\mu{g}/{\ell}$ , the accuracies compared to reference values were 97.0% and 92.6%, respectively. 2. Microwave digestion temperature($160~180^{\circ}C$), time(15~25min) and the ratio of urine/ashing acid(1:1~4:1) did not significantly affect lead absorbance and background absorbance. Therefore we set up a optimum analytical conditions as follows: temperature, $160^{\circ}C$; time, 15min; the ratio of urine/ashing acid, 4:1. after samples were digested by the above analytical conditions, lead absorbance and background absorbance was measured at $450^{\circ}C$ as ashing temperature. The average coefficient of variation (CV) of normal man was 12.4%, and lead worker was 6.2%. The average urine lead levels of standard urine were $10.66\mu{g}/{\ell}$ and $23.31\mu{g}/{\ell}$, the accuracies compared to reference values were 99.3% and 103.9%, respectively. From the results, we suggest that microwave digestion method is a more favorable method than matrix modifier addition method because of easiness to reduce organic matter, possibility to analysis at low temperature and accuracy.

Keywords

References

  1. The diagnosis of occupations Hunter, D.
  2. 예방의학과 공중보건 예방의학과 공중보건 편집위원회
  3. 금속과 사람 이영환;정문호
  4. 우리나라 주요 직업병 사례 노동부
  5. The biomonitoring program of the national center for environmental health NCEH(National Center for Environmental Health)
  6. Threshold limit valus (TLVs) for chemical substances and physical agent and biological exposure indices (BEIS) ACGIH(American Conference of Governmental Industrial Hygienists)
  7. Anal. Chem. v.65 no.5 Blood lead measurement takes the direct apporoach Novel, D.
  8. Arch. Environ. Health v.28 Laboratory diagnosis of increased lead absorption Baloh, R.W.
  9. New Engl. J. Med. v.289 Undue absorption of lead among children ; A new look at an old problem Lin Fu, J.S.
  10. Analytical Aspects and Health Significance v.Ⅰ Atomic absorption spectrometry in occupational and environmental health practice Tsalev D.L.;Zaprinov Z.K.
  11. Lead determination in blood and urine by AAS Stevens B.J.;Sanders J.B.;Stux R.
  12. Atomic Absorption Newsletter v.7 Olson A.D.;Hamlin W.B.
  13. 한국산업위생학회지 v.5 no.1 일부 산업 보건기관들의 혈중연 분석치 비교 안규동;이병국
  14. Application in Agriculture Biology and Medicine Atomic absorption spectroscopy Christian G.D.;Feldman J.J.
  15. Analyst v.106 Determination of lead in urine by atomic absorption spectroscopy with electrothermal atomisation Hodges D.J.;Skelding D.
  16. Atomic Spectroscopy v.6 no.5 Determination of urinary lead by electrothermal atomic absorption with the stabilized temperature platform furnace and martix modification Paschal D.C.;Kinberly M.M.
  17. Medical Laboratory Science v.46 Direct determination of lead in urine by flameless atomic absorption spectrophotometry with platform atomisation Mark Y.T.;Ho D.W.;Lam C.W.K.
  18. 한국의 산업의학 v.35 no.4 마이크로파 전처리법을 이용한 뇨중 연의 분석 노영만;한진구
  19. Clin. Chem. v.19 no.9 A simplified determination of urinary cadmium, lead, and thallium, with use of carbon rod atomization and atomic absorption spectrophotometry Kubasik N.P.;Volosin M.T.
  20. AAS, Clin. Chem. v.91 no.41 Microdetermination of Pb and Cd in blood and urine by graphite furnace AAS Allian P.;Mauras Y.
  21. Clin. Chem. v.25 Direct determination of Pb and Cd in whole blood and urine by flameless AAS Lagesson V.;Andrasko L.
  22. 예방의학회지 v.20 no.1 일부 중소도시 기혼여성의 혈액 및 뇨중 중금속 함량의 상관성에 관한 연구 황인담;기노석;이재형;박인서
  23. 대한화학회지 v.40 no.3 SR바탕보정장치와 D₂바탕보정장치가 부착된 GFAAS를 이용한 혈액중 카드뮴 분석 이석기;김풍작;정창웅
  24. Theory and Practical Introduction to microwave sample preparation Kingston H.M.;Jassie L.B.