Design of Neutron Shielder for Reducing Background of Low Level Gamma Ray Spectrometer

극저준위 감마선 분광시스템의 백그라운드 저감화를 위한 중성자 차폐체 설계

  • Published : 2001.06.30

Abstract

In order to shield the neutrons affecting the background of Low Level Gamma Ray Spectrometer, a neutron shielder was designed. The method used in this study for neutron shielding was the deceleration of fast neutrons by high density polyethylene(HDPE) and the absorption of those slowing-down neutrons by $B_4C$. The calculation results of neutron Interaction in HDPE using Monte Carlo simulation code MCNP4B showed that the thermal-neutron flux was maximum at 10 cm thickness of HDPE. The results also showed that 95% of the thermal neutrons were absorbed by 2 mm thickness of $B_4C$ absorber Consisted of 30 w% $B_4C$ and 70 w% polymer. The results of the Monte Carlo calculation were in good agreement with the experimental value obtained by a neutron shielding apparatus designed for this purpose.

극저준위 방사능측정시스템의 백그라운드에 영향을 주는 중성자를 차폐하기 위한 차폐체를 설계하였다. 중성자 차폐방법은 고 밀도 폴리에틸렌을 이용하여 고속중성자를 감속한 후 $B_4C$를 이용하여 감속된 열중성자를 흡수하는 방법을 이용하였다. 몬테카를로 모사방법인 MCNP4B 코드를 이용하여 계산한 결과 고 밀도 폴리에틸렌의 두께가 10 cm 일 때 열중성자속이 최대가 되는 것으로 나타났으며 감속된 중성자의 흡수는 용제에 자연상태의 $B_4C$ 분말을 30 w% 섞을 경우 2 mm의 두께에서 94%의 중성자 흡수가 일어나는 것으로 나타났다. 또한 몬테카를로 모사를 통한 계산결과의 타당성 여부를 조사하기 위하여 중성자 차폐실험 장치를 제작하여 실험 결과와 비교하였으며, 비교 결과 실험값과 일치하는 것으로 나타났다.

Keywords