The development of photo-diode dosimeter(PD-2000) for the diagnostic X-ray Energy

X선 진단영역 에너지 측정을 위한 Photo-Diode 선량계(PD-2000)의 개발

  • 김성철 (가천길대학 방사선과) ;
  • 이우철 (서울보건대학 의공학과) ;
  • 김정민 (고려대학교 보건대학 방사선과)
  • Published : 2000.12.30

Abstract

It was produced radiation dosimeter used photo-diodes for which ionization by x-ray was applied and evaluated the value of utility in clinics as compared with ion-chamber. The result obtained were as follows : 1. Comparison of ion-chamber with photo-diode dosimeter's x-ray output by the change of x-ray tube voltage, and the ratio of ion-chamber to diode was $0.96{\sim}1.02$ which was not affected by x-ray beam quality. 2. The ratio of ion-chamber to diode was 0.96 by change of tube current and 0.97 by change of exposure time that is not affected by x-ray quantity. 3. The ratio of ion-chamber to diode was $0.97{\sim}1.04$ by thickness and $0.93{\sim}1.10$ by radiation field that is little affected by second ray quantity. 4. Reproducibility of photo-diode dosimeter was 0.011(CV) and it is a good result. 5. Photo-diode dosimeter was affected by the surface angle of detector over 30 degrees. Produced dosimeter was small, light, and meets good result compared with ionization chamber. It was expected come into wide use in clinic.