Measurement of Absorbed Dose for High Energy Electron using $CaSO_4:Tm$ - PTFE TLD

고에너지 전자선의 흡수선량 측정에 있어서 TLD의 유용성

  • Published : 2000.06.30

Abstract

In this study, the highly sensitive $CaSO_4:Tm$-PTFE TLDs has been fabricated for the purpose of measurement of high energy electron. $CaSO_4:Tm$ phosphor powder was mixed with polytetrafluoroethylene(PTFE) powder and moulded in a disk type(diameter 8.5 mm. thickness $90\;mg/cm^2$) by cold pressing. The absorbed dose distribution and ranges for high energy electron were measured by using the $CaSO_4:Tm$-PTFE TLDs. The ranges determined were $R_{100}=14.5mm$, $R_{50}=24.1mm$ and $R_P=31.8mm$, respectively and the beam flatness, the variation of relative dose in 80% of the field size, was 4.5%. The fabricated $CaSO_4:Tm$-PTFE TLDs nay be utilized in radiation dosimetry for personal, absorbed dose and environmental monitoring.

Keywords