Abstract
The study was carried out to confirm the reliability of standard samples, low concentration $0.74mgN(NO_3)/l$, $0.58mgF^-/l$, high concentration $2.14mgN(NO_3)/l$, $1.25mgF^-/l$, on the base of quality control chart. The analytical results had the precise reliability with standard deviation of 0.010, 0.017, 0.056 and 0.019, respectively. When three persons A, B and C, who had different analytical experiences, analyzed the standard samples six times repeatedly, C person, who had more academic career and analytical experience, produced far better reliability for analytical results. Therefore, these standard samples were very suitable for external quality control test, because all persons' average values for analytical results were close to the values of prepared standard samples.
질산성질소와 불소의 저농도 $0.74mgN(NO_3)/l$, $0.58mgF^-/l$ 및 고농도 $2.14mgN(NO_3)/l$, $1.25mgF^-/l$ 표준시료에 대해 정도관리차트에 따라 신뢰도를 검증한 결과 정밀도가 각각 0.010, 0.017, 0.056, 0.019로 우수하게 나타났다. 분석의 숙련도가 다른 A, B, C 세명에 대해 각각의 표준시료로 6회 반복 측정분석한 결과 학력 및 연구경력이 가장 나은 C가 A, B보다 신뢰도가 우수한 측정분석결과를 산출하였다. 3명 모두 평균값이 조제값에 아주 가까운 분석데이타를 산출함으로서 이 표준시료를 외부정도관리 평가용으로 사용하기에 적합하다는 것을 알 수 있었다.