Abstract
Generally an operator would take notice at putting a patient under anesthesia. If the operation is executed in mistake, the patient is exposed to danger. The object of this Paper is that a system is developed for an accuracy of system and a convenience of user interface to prevent an operation of several elements of risk by mistake. The part of electrical system particularly is made for convenience of a manipulation using electrical switch and encoder. A real-time monitoring system is developed for an airway pressure and a gas concentration of carbon dioxide of patient using graphic LCD(liquid crystal display). Moreover, this flow control system could be developed control with accuracy by feedback control method. This is implemented using flow control valve and flow sensor. The implemented system gives convenience and precision of a manipulation of variable value using developed technique. This system shows guaranteed stabilization and confidence of anesthesia ventilator by notifying us that patient's state and information in case of being out of alarm range of variable value.
일반적으로 시술자가 환자에게 마취를 할 때에는 매우 주의하여야 한다. 만약 잘못 시행 될 때는 환자는 매우 위험한 상황에 빠지게 된다. 본 논문에서는 잘못 시술이 발생 될 수 있는 몇몇 위험요소들을 사전에 예방하기 위하여 시스템의 정밀성과 사용자의 편리성을 고려하여 구현하는 것을 목표로 하였다. 특히 시스템에서 전자적인 부분은 스위치와 엔코더를 이용하여 사용자 인터페이스에서 조작을 편리하게 하고, 그래픽 액정화면 표시기를 이용하여 환자의 기도압과 이산화탄소 파형을 실시간 모니터링 기능을 구현하였다. 또한 설정값의 정밀한 제어를 위해 기계적인 부분에서 유량 제어 밸브와 유량 센서를 이용하여 피드백 유량 제어 시스템을 구현하였다. 이러한 기술을 개발함으로써 시술자에게 설정치 조작의 편리성과 정확성을 가져다줄 뿐만 아니라 환자의 상태와 여러 가지 변수들의 허용 범위를 넘을 경우 정확하고 신속하게 정보를 알려줌으로서 마취기용 인공 호흡기의 안정성과 신뢰성이 확보될 수 있음을 알 수 있었다.