Optimal Design Method for an Actively Shielded MRI Superconducting Magnet

능동 차폐 MRI 초전도 자석에 대한 최적 설계 방법

  • Published : 2000.06.01

Abstract

This paper describes an optimal design method which is applied a weighted least square (WLS) method for Magnetic Resonance Imaging (MRI) system. An optimal design approach is presented for a homogeneity superconducting magnet with the superconducting active shield especially for a magnetic resonance imaging system. The WLS is used to obtain the optimal configurations using the least amount and minimum volume of conductor, exhibiting the smallest level of field inhomogeneity and resulting in the least level of stray field. The proposed model is used to design a multiple-shield configuration for a 1.5 T MRI magnet. The field homogeneity is required less than 5 gauss stray field contour within 4m axially and 3m radially from origin. The designed magnet with the actively magnetic shielding coil out of main coils is analyzed by FEM and theoretical analysis method, investigated the field homogeneity.

Keywords

References

  1. M.W.Garret , The Method of Zonal Harmonics in High Magnetic Field ,Willy, pp14-26 ,1961
  2. D.B.Monthomery, Solenoid Magnet Design , Robert E. Krieger Pub. Co., 1980
  3. W.Conley, Computer Optimization Techniques, rev.ed. Princeton, NJ , Petrocelli , 1984
  4. Alok Tripathi and T. S. Vedavathy, 'Electromagnetic Shielding Using Superconductors', Applied Superconductivity, Vol.2, No.1,pp.1-5, 1994
  5. Hironori Matsuba, 'Superconducting Magnetic Shield', Low Temperature Technology, Vol. 30, No.1, 1995
  6. A. Ishiyama and H. Hirooka, Magnetic Shielding for MRI Superconducting Magnets , IEEE Trans. Magn., Vol.27, No.2, pp. 1692-1695, 1991 https://doi.org/10.1109/20.133514
  7. 戶田 博章, '磁氣 遮蔽', 一本 高壓カス工業株式會社 企劃室, 1990
  8. 非線型 最適化 プログラシソグ, ASNOP 硏究會, 日刊工業新聞社, 1991