Abstract

The purpose of study was to measure the interrater reliability of the newly developed caries diagnostic instrument, Diagnodent(KaVo, Germany), using a new laser fluorescence techinique. One hundred and three extracted human premolars and molars were examined by Diagnodent, three times for fissures and five times for smooth surfaces. There was no significant difference among the three fissure means, but two combinations showed significant differences among the five smooth surface means. Correlation coefficients of Pearson and Spearman were all above 0.9(P<0.01). The $\alpha$ values from reliability analysis were 0.9980 or 0.9981 for fissures and 0.9992 for smooth surfaces.

연구목적은 새로운 레이저형광측정기술을 응용한 휴대용 우식진단기구인 Diagnodent를 사용한 우식진단법의 검사자내 신뢰도를 분석하는 것이었다. 발거된 사람 대구치와 소구치를 대상으로 열구를 3회, 평활면을 5회 반복 검사하였다. 측정치들의 평균치간 차이의 유의성을 t-검사로 검정하였고 측정치들간의 상관성을 Pearson 상관계수와 Spearman 순위상관계수로 산출하였으며 척도 신뢰도분석으로 검사자내 신뢰도를 분석하였다. 연구성적에서, 열구측정치들 간에 유의한 차이는 없었으나, 평활면측정치들간에는 10개 조합 중 2개에서 유의한 차이가 있었다(P<0.05). 그러나, Pearson 상관과 Spearman 순위상관은 모두 상관계수 0.9 이상의 유의한 상관성을 나타내었고(P<0.01), 신뢰도의 값은 열구측정에서 0.9980 내지 0.9981로, 평활면측정에서 0.9992로 산출되었다. Diagnodent를 사용한 우식진단법은 생체외에서 한 검사자가 반복 검사를 시행하였을 때 완전한 일치에 가까운 재현도, 곧 매우 높은 검사자내 신뢰도를 가지고 있는 것으로 나타났다.