Development of Prototype Electronic Dosimeter using the Silicon PIN Diode Detector

실리콘 PIN 다이오드 검출기를 이용한 전자선량계 개발

  • Lee, B.J. (Korea Atomic Energy Research Institute) ;
  • Kim, B.H. (Korea Atomic Energy Research Institute) ;
  • Chang, S.Y. (Korea Atomic Energy Research Institute) ;
  • Kim, J.S. (Korea Atomic Energy Research Institute)
  • Published : 2000.12.30

Abstract

A prototype electronic dosimeter(PED) adopting a silicon PIN diode detector as a radiation detector has been developed, manufactured and test-evaluated. A radiation signal processing circuit has been electronically tested and then the radiation detection characteristics of this PED has been performance-tested by using a reference photon radiation field. As a result in a electronic performance test, radiation signals from a detector were well observed in the signal processing circuit. The radiation detection sensitivity of this PED after several test-irradiations to $^{137}Cs$ gamma radiation source appeared to be 1.85 cps/$Gy{\cdot}h^{-1}$ with 19.3% of the coefficient of variation, which satisfied the performance criteria for the active personnel radiation monitor. Further improvement of the electronic circuit and operating program will enable the PED to be used in personal monitoring purpose.

실리콘 PIN 다이오드 검출기를 이용하여 전자선량계(PED)를 설계 및 제작하고, 전자신호회로의 성능과 방사선 검출 반응특성을 평가하였다. 전자회로의 성능검사에서는 방사선에 의한 검출기의 신호가 신호처리회로에서 양호하게 출력되었으며, $^{137}Cs$ 감마 방사선을 이용한 방사선조사 시험결과 선량계의 반응도 및 분산계수는 각각 1.85 cps/$Gy{\cdot}h^{-1}$ 및 19.3%로 측정되어 전자선량제의 성능기준을 만족하는 것으로 나타났다. 본 전자선량계의 시작품은 향후 전자회로의 성능향상과 작동 프로그램을 보완 및 개선함으로써 개인 방사선 모니터링에 활용될 것으로 기대된다.

Keywords