Abstract
Triple points of high purity materials have been used to calibrate primarily the capsule-type platinum resistance thermometer (PRT) in the temperature range of the triple point of equilibrium hydrogen (13.8033 K) and water (273.16 K). In this work, triple points of Ne,
평형수소 삼중점 (13.8033 K)부터 물의 삼중점 (273.16 K) 온도영역에서 표준온도계로 사용되는 캡슐형 백금저항온도계를 일차 교정하려면 고순도 물질의 삼중점이 사용된다. 국제온도눈금-1990 (ITS-90)을 확립하기 위해 평형수소, 네온, 산소, 아르곤, 수은 및 물의 삼중점 중에서 평형수소의 삼중점을 제외한 5 개의 삼중점을 실현하였다. 각각의 삼중점에서 국제비교용으로 채택된 캡슐형 백금저항온도계 2대의 저항을 2회 반복하여 측정하였다. 측정된 삼중점에서의 합성표준불확도는 A형 불확도와 B형 불확도 평가방법을 고려하여 산출하였다. 한국표준과학연구원에서 ITS-90에 정의된 네온, 산소, 아르곤, 수은 및 물의 삼중점에서의 합성표준불확도 추정결과는 각각 0.18 mK, 0.14 mK, 0.14 mK, 0.24 mK 및 0.11 mK 이었다.